Regular Maintenance
19
Step 3 Place a tag on the WRO 300, stating that chemical disinfec-
tion is in progress and what type of disinfectant is being used.
Step 4 Insert the wand connector into the chemical intake port of the
WRO 300 (see the illustration below), press firmly and turn
the connector downwards to lock. Check that the connector is
securely in place.
Step 5 CHEM SELECT is shown in the display and flashes.
Step 6 Insert the open end of the disinfection wand into a container
with sufficient amount of disinfectant (refer to the Approved
Chemical Disinfectants on page 17). Check that the open end
of the intake wand is below the solution level and that the
clamp is open. The container must be below the level of the
chemical intake port to prevent siphoning chemical into the
unit.
Disconnect the quick connector
Nipple at the Dialysis machine
WRO 300
Insert the connector and press Turn to lock
Disinfection Wand Connector
Container
Container
Disinfection Wand
Connector
Correct
Wrong