Step 7 From CHEM SELECT, press to enter the list of available
Cleaning and chemical disinfection programs.
Step 8 Select preferred chemical disinfection program by using the
arrow buttons.
Step 9 Initiate the disinfectant intake by pressing until light goes
on.
The table below explains how to pause, interrupt or restart the proce-
dure.
Step 10 401 REMOVE WAND is displayed and the buzzer sounds and
is flashing, press .
NOTE
The display shows present phase of the chemical disinfection proce-
dure.
NOTE
The container can be empty before the intake phase is finished.
CAUTION
The chemical disinfection has to be restarted if alarm 406
INSUFFICIENT INTAKE has appeared.
Action
Pause of chemical
intake phase
Press until light flashes. To continue the intake
press again until light goes on.
Interrupt chemical
disinfection procedure
Disconnect the wand connector and press until
light goes on, to initiate rinse.
Restart of intake phase
Remove the wand connector and insert it again CHEM
INTAKE RESTART appears on the display and
flashes. Then press until light goes on.
MINNCARE
CHEM INTAKE
X