Operator`s manual

WRO 300
P/N 3027435 Rev. C
Regular Maintenance
23
3.2 Cleaning
3.2.1 Acid Cleaning
Acid cleaning should be performed when:
Rejection rate has decreased by more than 5 percent from initial
value.
Sufficient output flow is not obtained.
As preventive measure when a softener is not used.
As a method to remove transitional metals from the RO membrane
prior to chemical disinfection.
Dissolve 100 ml of citric acid in 200 ml of product water from the
WRO 300. This will result in a 2 percent solution in the WRO 300 or
use MINNCLEAN
®
AC according to manufacture’s instruction.
Refer to Cleaning Procedure on page 23.
3.2.2 Alkaline Cleaning
Alkaline cleaning should be performed when:
Product performance is affected and cleaning with acid does not
improve performance.
Organic fouling is suspected.
Use MINNCLEAN
®
TF according to manufacture’s instruction.
Refer to Cleaning Procedure below.
3.2.3 Cleaning Procedure
Step 1 Press until the display shows RUN. Wait until the product
water conductivity stabilizes and record the value. Press
until light goes out to stop the WRO.
NOTE
During the intake phase of the central chem program, the product
water loop will not be pressurized until the correct conductivity has
been obtained, which will take a few minutes.
NOTE
Other acid membrane cleaner may be used as an alternative cleaner
for the WRO 300. Follow manufacturer’s guidelines to prepare the
cleaning solution.