Operator`s manual

WRO 300
P/N 3027435 Rev. C
Regular Maintenance
25
3.3.1.2 Preservation Procedure
Step 1 Follow the chemical disinfection procedure (refer to Chemical
Disinfection on page 17) Step 2 through Step 12, but instead
select MEMSTOR in the CHEM SELECT menu.
Step 2 After Step 12: the display will show 404 RINSE REQUIRED
and the buzzer will sounds and the indications and
flash. Press , or to acknowledge the alarm — the
alarm list unlocks.
Step 3 Turn off the WRO 300 using the mains power switch behind
the detachable cover.
3.3.2 Rinse After Preservation
Step 1 Turn on the WRO 300 using the mains power switch behind
the detachable cover.
Step 2 The indicators and flash and the display shows 404
RINSE REQUIRED.
Step 3 Press until light goes on to initiate rinse.
Step 4 When the rinse program is complete, the display will show
PERFORM RESIDUAL TEST. The indication remains
lighted and the flashes.
3.3.3 Check for Complete Rinse-out
The required rinse program in the WRO has been designed and vali-
dated to ensure complete rinse-out of the preservation.
Step 1 Press until light goes on to start the WRO—
indicator lights and flashes.
Step 2 Let the WRO run in normal operation for at least 5 minutes.
Then press or to show the conductivity value in the
display (within 15 seconds the display returns to PERFORM
RESIDUAL TEST). Check and verify that the product water
conductivity remains stable and constant.
CAUTION
If the conductivity limit is not achieved, alarm 406 INSUFFI-
CIENT INTAKE will appear and the Preservation procedure
has to be restarted.