Operator`s manual
P/N 3027435 Rev. C
Regular Maintenance
26
WRO 300
Step 3 Confirm by pressing until light goes out. Indication
remains lighted.
Step 4 The WRO is now ready to use.
3.4 Flush
3.4.1 Manual Flush
This mode provides a short flush of the WRO at elevated flow rate (5
minutes preset).
Step 1 Press until SELECT appears.
Step 2 Scroll with to select MANUAL FLUSH.
Step 3 Press to initiate manual flush (MANUAL FLUSH shows
and lights).
3.4.2 Auto Flush
During standby auto flush is regularly performed to exchange water in
the WRO, if it is enabled in preset. At specified intervals the water cir-
culates in the WRO and some water goes to drain.
3.5 Exterior Cleaning
Wipe the outside of the WRO 300 with a cloth moistened with ethanol
(70 percent) or isopropanol (60 percent).
NOTE
Do not use iodine-based or tenside-containing disinfectants as these
may crack or discolor the plastic materials.