Operator`s manual
WRO 300
P/N 3027435 Rev. C
Troubleshooting
41
5.2 Boot loader
This alarm indicate that the software is corrupt.
408 INSERT WAND WRO 300 WRO 300 stops, buzzer sounds with alarm
sound,
flashes, message text displays.
Reason A central chemical disinfection has been initi-
ated via the remote control and the disinfec-
tion wand is not inserted.
Action
1 Press ,
or to silence buzzer.
2 Insert the disinfection wand to continue the
central chemical disinfection procedure.
No.
Alarm and
notifications Definition
LOAD APPLICA-
TION FAILED
WRO 300 LOAD APPLICATION FAILED is
shown on the display.
Reason Software malfunction.
Action Notify technical personnel.
No.
Alarm and