Operator`s manual
P/N 3027436 Rev. C
Introduction
12
WRO 300 H
adjusted at installation to a lower value by a qualified technician to fit
local conditions, especially in cases of low feed water conductivity
(<100 µS/cm). The limit should be set at least 5 percent lower than the
initial value recorded at installation.
The low rejection rate notification will be displayed when the
WRO 300 H enters standby if the set limit has been exceeded for more
than 5 consecutive minutes during the previous run period.
1.5 Mains Power Switch
The mains power switch is only used in service situations to turn off the
power. The mains power switch is located on the rear side behind a
detachable cover.
1.6 Operator Panel
1.6.1 Description of Buttons
Buttons Description
Press to start operation (RUN).
Press to stop operation (STANDBY).
Press to stop heat disinfection procedure and low flow heat.
Press to enter
SELECT (only in Standby mode)
Press to start the selected procedure (SELECT or CHEM SELECT).
Press during the chemical intake procedure to pause.
Press during dwell period to initiate rinse.
Press to silence audible alarms.
Press to unlock the display and enter the alarm list.
Mains power switch