Operator`s manual

WRO 300 H
P/N 3027436 Rev. C
Operation
15
2 Operation
2.1 Start
Press until light goes on.
2.2 Stop
Press until light goes out and the WRO 300 H goes to standby
mode.
WARNING
This device does not remove chlorine and chloramines. If
these substances are present in the feed water, carbon filtra-
tion is required to remove these substances. Severe patient
injury may otherwise occur. A test for total chlorine of the pre-
treated water must be performed prior to initiating dialysis
treatment. The level of total chlorine must be below 0.1 mg/l
(ppm).
NOTE
will light up and the WRO 300 H will go into active cooldown if the
internal water temperature exceeds 45 °C ( will still be lit). The pro-
cedure will continue until the temprature is below 45 °C.
NOTE
If power failure occurs during operation mode the procedure contin-
ues in the same phase as when the power failure occurred when the
WRO 300 H starts up again.
RUN
X
μS
/
cm
STANDBY