Regular Maintenance
17
3 Regular Maintenance
3.1 Heat Disinfection
The required heat disinfection frequency to fulfil the desired microbio-
logical requirements for the product water depends on several factors
such as:
• the quality of the feed water
• local regulations regarding the microbiological quality of dialysis
water, etc.
No general rules can be given that cover all situations. The disinfection
schedule should be based upon microbiological testing performed by
the clinic. A maximum period between disinfections should then be
established to ensure acceptable bacteriological quality of the product
water according to the clinic’s standards.
MCP recommends a minimum frequency of weekly heat disinfection to
ensure consistent microbiological quality of the product water.
If the WRO 300 H will not be used for an extended period of time, reg-
ular automated heat disinfection preset by the time channels will main-
tain the microbiological quality of the product water.
3.1.1 Heat Disinfection Procedure
The following paragraphs describe the heat disinfection of the WRO
300 H and the product water loop.
NOTE
The procedure below assume factory settings of the protocol.
NOTE
If the internal water temperature exceeds 45 °C when initiating the
preferred procedure the display reverts to the
SELECT menu or the
CHEM SELECT menu.
This situation is resolved by initiating RUN and letting the WRO 300 H
operate and perform an active cooldown until the temperature is
below 45 °C.
NOTE
The dialysis machine shall be turned off when the heat disinfection
procedure is performed.
NOTE
The heat disinfection procedure can also be initiated by the time chan-
nels. Further information of time channel settings refer to the Service
Manual.