Operator`s manual
WRO 300 H
P/N 3027436 Rev. C
Regular Maintenance
19
3.2.1 Approved Chemical Disinfectants
•MINNCARE
®
Cold Sterilant
Required amount: 50 ml of concentrated disinfectant
3.2.2 Test Kit Example
• Minncare Residual Test Strips, MCP order number 185-40-004
(preferred).
• STERICHEK
®
Residual Peroxide, MCP order number WT811905
(alternate).
3.2.3 Chemical Disinfection Procedure
Chemical disinfection of the WRO, including product water loop.
Step 1 Place the WRO into standby mode.
Step 2 Disconnect the WRO 300 H from the dialysis machine with
the quick connector on the product water loop. Alternatively,
CAUTION
Disinfectants may be toxic. Refer to the manufacturer instruc-
tions.
CAUTION
A test for residual disinfectant after rinse must be performed
before the initiation of the next dialysis session. The residual
concentration of the disinfectant in the fluid must be below lev-
els specified by the clinic or by national standards. It is essen-
tial to use an appropriate test method, either with proven
sensitivity for the disinfectant or recommended by the manu-
facturer of the disinfectant.
CAUTION
The chemical container must be located below the chemical
intake.
CAUTION
When requested by the WRO unit, remove the disinfection wand
from the chemical intake. Failure to remove the disinfection
wand can cause siphoning of the chemical solution from the bot-
tle resulting in the tank overflowing chemical solution through the
back of the machine.
NOTE
If power failure occurs during chemical disinfection 404 RINSE
REQUIRED is displayed when the WRO 300 H restarts (also 401
REMOVE WAND is displayed during the intake phase). Press until
light goes on, to initiate rinse.