Regular Maintenance
20
WRO 300 H
the connector can be left in place. Refer to the caution mes-
sage below.
Step 3 Place a tag on the WRO 300 H, stating that chemical disin-
fection is in progress and what type of disinfectant is being
used.
Step 4 Insert the wand connector into the chemical intake port of the
WRO 300 H (see figure below), press firmly and turn the con-
nector downwards to lock. Check that the connector is
securely in place.
Step 5 CHEM SELECT is shown in the display and starts to flash.
CAUTION
If the WRO 300 H is not disconnected from the dialysis
machine during the chemical disinfection procedure, the test
for residuals after the chemical disinfection procedure must be
performed on the dialysis machine according to the Operator’s
Manual of this device.
Disconnect the quick connector
Nipple at the Dialysis machine
WRO 300 H
Insert the connector and press Turn to lock