Regular Maintenance
21
Step 6 Insert the open end of the disinfection wand into a container
with sufficient amount of disinfectant (refer to Approved
Chemical Disinfectants on page 19). Check that the open end
of the intake wand is below the solution level and that the
clamp is open. The container must be below the level of the
chemical intake port to prevent siphoning chemical into the
WRO 300 H unit.
Step 7 From CHEM SELECT, press
to enter the list of available
cleaning and chemical disinfection programs.
Step 8 Select preferred chemical disinfection program by using the
arrow buttons.
Step 9 Initiate the disinfectant intake by pressing until light goes
on.
NOTE
The display shows present phase of the chemical disinfection proce-
dure.
NOTE
The container can be empty before the intake phase is finished.
CAUTION
The chemical disinfection has to be restarted if alarm 406
INSUFFICIENT INTAKE has appeared.
Container
Disinfection Wand
Connector
Correct
Wrong
Disinfection Wand Connector
Container
MINNCARE
CHEM INTAKE
X
μS
/
cm