Regular Maintenance
24
WRO 300 H
3.2.5 Central Chem
The central chem disinfects the WRO and also sends disinfectant into
the dialysis machine. For more detailed information on how to perform
central chem together with Gambro
®
dialysis machines, please contact
Technical Service.
Follow the Chemical Disinfection Procedure on page 19, but start at
Step 3 and follow the instruction through Step 13.
In Step 8 select the preferred central chem protocol in the CHEM
SELECT menu.
3.3 Cleaning
3.3.1 Warm Acid Cleaning
Warm acid cleaning should be performed when:
• Rejection rate has decreased by more than 5 percent from initial
value.
• Sufficient output flow is not obtained.
• As preventive measure when a softener is not used.
• As a method to remove transitional metals from the RO membrane
prior to chemical disinfection.
CAUTION
If the produced disinfection solution is fed into a device other
than Gambro
®
dialysis machines, it is the user’s responsibility
to ensure that the connected device is compatible with perace-
tic acid disinfectants and to establish the procedure for the
WRO in combination with the connected device.
CAUTION
As the disinfection solution produced by the WRO will enter
any device connected to the quick connector on the product
water loop, residual test after disinfection must be performed
on the connected device in accordance with the Operator’s
Manual for the connected device.
NOTE
During the intake phase of the central chem program, the product
water loop will not be pressurized until the correct conductivity has
been obtained, which will take a few minutes.