Regular Maintenance
26
WRO 300 H
Step 4 Confirm by pressing until light goes out. Indication
remains lighted.
3.4 Long Term Storage
If the WRO 300 H will not be used for an extended period of time, reg-
ular automated heat disinfection preset by the time channels, will main-
tain the microbiological quality in the WRO.
3.4.1 Preservation
If automated heat disinfection is not possible to perform when the
WRO 300 H is taken out of operation for an extended period of time, a
chemical preservation must be done. For example, perform preserva-
tion if electrical and water connections are disconnected.
3.4.1.1 Preservation agents
Use MEMSTOR and dissolve 100 ml in two liters of potable tap water.
Warm water (35-40°C) speeds up dissolution.
3.4.1.2 Preservation Procedure
Step 1 Follow the chemical disinfection procedure (refer to Chemical
Disinfection Procedure on page 19) Step 2 to Step 12, but
instead select MEMSTOR in the CHEM SELECT menu.
Step 2 After Step 12: the display will show 404 RINSE REQUIRED
and the buzzer will now sound and the indications and
will flash. Press , or acknowledge the alarm
and the alarm list will be unlocked.
Step 3 Switch off the WRO 300 H with the mains power switch
behind the detachable cover.
3.4.2 Rinse After Preservation
Step 1 Turn on the WRO 300 H with the mains power switch behind
the detachable cover.
Step 2 The indication and flashes and the display shows
404 RINSE REQUIRED.
Step 3 Press until light goes on to initiate rinse.
CAUTION
If the conductivity limit is not achieved, alarm 406 INSUFFI-
CIENT INTAKE will appear and the preservation procedure has
to be restarted.
RINSE
hh:mm