Operator`s manual
WRO 300 H
P/N 3027436 Rev. C
Regular Maintenance
27
Step 4 When the rinse program is completed the display will show
PERFORM RESIDUAL TEST. The indication will remain
lighted and the will flash.
3.4.3 Check for Complete Rinse-out
The required rinse program in the WRO has been designed and vali-
dated to ensure complete rinse-out of the preservation.
Step 1 Press until light goes on, to start the WRO.
Indication will light and will flash.
Step 2 Let the WRO run in normal operation for at least 5 minutes.
Then press or to show the conductivity value in the
display (within 15 seconds the display will return to PERFORM
RESIDUAL TEST). Check and verify that the product water
conductivity remains stable and constant.
Step 3 Confirm by pressing until light goes out. Indication
remains lighted.
Step 4 The WRO is now ready to use.
3.5 Flush
3.5.1 Manual Flush
This mode provides a short flush of the WRO at elevated flow rate (5
minutes preset).
Step 1 Press until SELECT appears
Step 2 Scroll with to select MANUAL FLUSH.
Step 3 Press to initiate Manual Flush (MANUAL FLUSH will be
displayed and will light).
3.5.2 Auto Flush
During standby, auto flush is regularly performed to exchange water in
the WRO, if enabled in preset. At specified intervals the water circu-
lates in the WRO and some water goes to drain.
3.6 Exterior Cleaning
Wipe the outside of the WRO 300 H with a cloth moistened with etha-
nol (70 percent) or isopropanol (60 percent).
NOTE
Do not use iodine-based or tenside-containing disinfectants as these
may crack or discolor the plastic materials.