Operator`s manual
P/N 3027436 Rev. C
Technical Data
32
WRO 300 H
Power supply
Mains Voltage 100 to 240 VAC (line voltage range 85
to 264 VAC)
Frequency 50 or 60 Hz (line frequency range 45
to 65 Hz)
Power Consumption
Max
1500 W (100 VAC)
1380 W (115 VAC)
1850 W (220-240 VAC)
Appliance inlet According to IEC 60320; C14
Cable Cable with protective earth
Conductor length max. 3.5 m
minimum 1 mm² (220-240 VAC)
AWG 16 at 100 and 115 VAC
Mains plug North America type—100 and 115
VAC, Hospital grade, earthed plug,
type IEC 60 083; A5-15
Fuses 15 A slo blo (115 VAC)
15 A slo blo (100 VAC)
Earth Leakage Current Max. 250 µA (220–240 VAC)
Max. 145 µA (115 VAC)
Max. 140 µA (100 VAC)
Connection of
external
equipment
External connector 8 pins REDEL, yellow.
Not used.
External connector 8 pins REDEL, black. RS-232
Logging interface
Max input voltage ±15 VDC
High level min
output voltage
5.0 VDC
Low level max
output voltage
5.0 VDC
Max Current ±5 mA DC
External connector 8 pins REDEL, green.
Not used.
Sound
Sound power level Less than 65 dBa during normal oper-
ation
Values










