Operator`s manual
WRO 300 H
P/N 3027436 Rev. C
Technical Data
33
4.2 Chemical Disinfection and Cleaning
4.3 Physical Data
Disinfection
The following disinfectants may be administered via the chemical
intake.
• MINNCARE
®
Cold Sterilant
• Other peracetic acid disinfectants, provided that they are
approved by the manufacturer for disinfection of thin film
composite reverse osmosis membranes made of modified
polyamide. Follow the manufacturer´s instructions for Use for
the specific disinfectant.
Cleaning
agents
The following cleaning agents may be administered via the
chemical intake:
• Citric acid
•Minnclean® AC
•Minnclean® TF
•Acetic Acid (5%)
• Other cleaning agents may also be used, provided that they
are approved by the manufacturer for cleaning of thin film
composite reverse osmosis membranes made of modified
polyamide. Follow the manufacturer's instructions for use for
the specific cleaning agent.
Preservation
The following preservations may be administered via the chemi-
cal intake and feed water inlet:
• MEMSTOR ®
• MEMSTOR ® in combination with 9 % glycerol
•Formalin
Exterior
cleaning
• Ethanol (70%)
• Isopropanol (60%).
Measurements Values
Weight
33 kg (72.8 lb)
Height
563 mm (22.2 inch)
Depth
Max 520 mm (20.5 inch)
Footprint 380 mm (15.0 inch)
Width
Max 205 mm (8.1 inch)
Footprint 185 mm (7.3 inch)
Internal fluid
volume
Approximately 3.5 litres excluding the product water loop