Operator`s manual

P/N 3027436 Rev. C
Technical Data
34
WRO 300 H
4.4 Materials Contacting Product Water
4.5 Environmental Data
Type Material
Polymers
PA (Polyamide)
PEEK (Polyetheretherketone)
PEI (Polyetherimide)
PE (Polyethylene)
PFA (Fluorocarbon)
POM (Polyoxymethylene)
PP (Polypropylene)
PPS (Polyphenylensulphite)
PVC (Polyvinylchloride)
PVDF (Polyvinyliden fluorid)
Rubber
Q (Silicone)
Metals
Titanium
Stainless steel
Others
Ceramic
Value
Operation
Ambient Temperature range +10 to +40 °C
Relative Humidity range 30 to 85% RH
Air Pressure range 700 to 1060 hPa
Transportation
and storage
Ambient Temperature range -10 to +40 °C
Relative Humidity range 10 to 95% RH
Air Pressure range 500 to 1060 hPa
For transportation and storage below the freezing point, the
WRO 300 H must be filled with a preservation solution of 2%
MEMSTOR, at least 9% glycerol and the remaining parts RO
water.
During transportation and storage the equipment must be kept
in its original packing. If transportation or storage time is more
than 15 weeks, the environmental data relating to the operation
must be followed.
If condensation occurs when moving the equipment between
locations with different temperatures and high relative humidity
(for example. outdoor and indoor locations), the inside of the
equipment must be allowed to dry before switching on the
equipment.