BetaGauge II™ Pressure Calibrator User Guide
Edition 2.0 (May 2006) Copyright © 2006, Martel Electronics Corporation All Rights Reserved. This documentation is the confidential and proprietary intellectual property of Martel Electronics Corporation. Any unauthorized use, reproduction, preparation of derivative works, performance or display of this document, hardware, or software represented by this document, without the express written permission of Martel Electronics Corporation, is prohibited.
Table of Contents Table of Contents .............................................. iii About this Guide...................................................... xi Comments ...............................................................xii Factory Sales.......................................................... xiii 1- General Description..................................... 1-1 Overview................................................................ 1-1 Carrying Case ...........................................
3- Basic Operations ......................................... 3-1 Overview................................................................ 3-1 Selecting a Channel for Specific Functions ........... 3-2 Procedure ........................................................... 3-2 Selecting Pre-Programmed Display Units.............. 3-3 BetaPort–DC Module ........................................ 3-3 BetaPort–P Pressure Module ............................. 3-4 Selecting a Water Column Measurement .......
Setting LCD Preferences........................................ 4-4 Setting the Backlight Off Delay Time................ 4-4 Procedure ....................................................... 4-4 Setting the LCD Contrast................................... 4-5 Procedure ....................................................... 4-5 Recovering the LCD Contrast............................ 4-6 Setting the Automatic Shut Off Delay Time .......... 4-7 Procedure ........................................................
- Application Notes ........................................6-1 Overview................................................................ 6-1 Modes of Operation ............................................... 6-1 Other Necessary Equipment................................... 6-2 Calibration Tests.................................................... 6-3 Electrical to Pressure Test ..................................... 6-4 Conventional Mode ........................................... 6-4 SYS Mode ...............
Method 2......................................................6 -52 Temperature Indicator or Loop Test ....................6 -54 SYS Mode ....................................................... 6 -54 7- Documenting Calibrator (SYS) Mode ........7-1 Overview................................................................ 7-1 Acquiring Instrument Maintenance Software ........ 7-1 Before You Begin.................................................. 7-2 Downloading Instrument Tests ..........................
Specifying Post-Test Status Information......
A- Battery Care/Maintenance ..........................A-1 Overview...............................................................A-1 Battery Packs ........................................................ A-4 Guidelines for Using Batteries.............................. A-5 Installing a Battery Pack ....................................... A-5 Removing a Battery Pack......................................A-6 Viewing the Battery Voltage................................. A-7 Charging the Batteries ................
Power Supply........................................................ B-6 Circuit Protection.................................................. B-6 Pressure Connections............................................ B-6 Electrical Connections .......................................... B-6 Weight .................................................................. B-6 Size ....................................................................... B-6 C- Conversion Factors ....................................
xi
Introduction The BetaGauge II™ Pressure Calibrator is the most advanced and accurate handheld pressure calibrator available. This calibrator was engineered to perform in any industrial environment and is equipped with everything you need for maintaining your pressure instruments to your manufacturer’s specifications. About this Guide This guide provides complete information on the proper use and operation of the BetaGauge II.
Section 7 explains how to use the BetaGauge II with third party software to provide automated computer aided calibration. Appendix A provides the dangers, warnings, and cautions for using battery packs and explains other battery related information. Appendix B lists the detailed specifications for the BetaGauge II. Appendix C lists the pressure conversion factors for converting pounds per square inch (PSI) to other pressure units. Appendix D describes the necessary service required for the BetaGauge II.
Comments Martel Electronics Corporation is interested in your response to our products and to this guide. If you have any comments or suggestions, you may contact us as follows: Mail: Martel Electronics Corporation 1F Commons Drive, #39 P.O. Box 770 Londonderry, NH 03053 Voice: Inside the United States 1-800-821-0023 Outside the United States 1-603-434-1433 E-Mail: sales@martelcorp.com Please tell us how to contact you if we have any questions or follow-up to your inquiry.
xiv
1- General Description Overview The BetaGauge II pressure calibrator is the most advanced and accurate handheld pressure calibrator available today. Its 3½ inch diagonal liquid crystal display (LCD), membrane keypad, and user-prompt programming make the BetaGauge II easy to operate.
Carrying Case The BetaGauge II comes with a two-section padded vinyl carrying case. It is recommended that you use the case at all times to protect the calibrator from damage due to chemicals, dust, and dirt. The carrying case is described below: Front Section. The calibrator is secured in the front section of the case with an elastic strap that fits between the keypad and the display. Back Section.
Standard Equipment The BetaGauge II is supplied with the following standard equipment: One pressure calibrator hand held (HHU) unit One standard voltage/current DC input module One battery pack (nickel-cadmium, rechargeable) One set of test leads One battery charger One vinyl carrying case One 9-pin D shell to 9-pin D shell serial cable Optional Equipment The BetaGauge II is available with any of the following options: Additional standard input modules (28 pressure ranges to choose from)
BetaGauge II LCD The BetaGauge II liquid crystal display is divided into four sections. These sections are described below in the order that they appear on the display from top to bottom. Refer to the illustration below: Figure 1-2 BetaGauge II LCD Time/Date/Battery Indicator. The calibrator’s current time and date appears to the left of this section and the approximate battery voltage indicator appears to the right. Channel 1 Input Information.
Keypad Description The keypad on the BetaGauge II allows you to enter data, scroll through the menus, and select menu items. A sample keypad is shown below: Figure 1-3 BetaGauge II Keypad The BetaGauge II keypad consists of the following key groups: Channel keys Function keys Number keys Special keys Each of the key groups are described on the following pages.
Channel Keys The channel keys (CH1 and CH2), in the upper left and right corners of the keypad, allow you to select Channel 1 or Channel 2 for performing specific configuration or modification functions, such as damping or zeroing a module. Channel 1 data appears in the top half of the LCD display; Channel 2 data appears in the bottom half. Function Keys The function keys (F1 through F4), across the top of the keypad, allow you to select specific menu items.
Special Keys The special keys, in the right portion of the keypad, allow you to perform specific BetaGauge II operations. These special keys are described below: Key Allows you to … MENU Scroll forward through the menu items under the Setup menu and the Calibration menu. ON/OFF Turn the BetaGauge II on and off. ZERO Set any displayed value to zero. (Starburst) Toggle a display backlight on and off as needed. Refer to Section 4 for information on how to set the time delay for the backlight shut off.
Menus The BetaGauge II Calibrator has two main menus: Setup menu Calib menu See the BetaGauge II menu structure below: BetaGauge II Main Menu Structure Top Level Sub Menus (Shaded) Program Selections (not shaded) Setup Menu 1 Trip Unit Damp Dis. Enab Reset 0.0s 2.0s 5.0s 10.0s 1 V/mA Sys Owner Beep (MasterCAL IMMS (SYS) Mode) Dis. Enab LCD LED Contrast Dis.
Setup Menu The Setup menu contains the following menu items: Menu Item Allows you to … Trip Detect trips from either wet or dry contacts using an installed DC module. Refer to Section 3 for more information. Unit Select the desired pressure units for display when using a BetaPort–P pressure module. Refer to Sections 3, 4, and 5 for more information. Damp Select the damping factor for fluctuating inputs. Refer to Section 3 for more information.
Menu Item Allows you to … Mod View module specific information such as serial number, range, and last calibration date. Refer to Section 3 for more information. Time Set the time. Refer to Section 4 for more information. Date Set the date. Refer to Section 4 for more information. Format Set the format for the time and date. Refer to Section 4 for more information. Mode Set up the calibrator for using advanced display modes such as user-defined pressure units, percent scale, percent error, etc.
Calib Menu The Calib Menu contains the following menu items: Menu Item Allows you to … Main (Reserved for factory use.) Info Specify details on who owns the calibrator and view the version of firmware currently installed. Refer to Section 4 for more information. Accessing the Menu Items The menus and menu items appear at the bottom of the LCD.
General Description 1-12
2- Input Modules Overview The BetaGauge II is a two-channel, multi-range instrument. You can change ranges simply by plugging in new input modules, and you may install or remove modules even while the unit is powered up. The calibrator automatically recognizes the module types and displays the appropriate units.
Installing and Removing Modules You may insert or remove a module while the calibrator is on or off. If the calibrator is on, the module will selfinitialize within 3 seconds and a reading will display on the LCD. When facing the calibrator, the left module bay corresponds to Channel 1, and the right module bay corresponds to Channel 2. CAUTION: Use care when handling modules. It is recommended that you store unused modules in the carrying case supplied with the calibrator.
Removal Procedure To remove a module from the BetaGauge II calibrator, place one hand on the calibrator and use the other hand to firmly grasp the module and pull directly out from the calibrator. NOTE: If a BetaPort–P pressure module is removed or disconnected while the BetaGauge II is reading pressure units, the display will show an invalid pressure reading.
Procedure Use the following procedure to connect the BetaPort– DC electrical module to a field device: NOTE: Install the module in the calibrator prior to connecting the module to field devices. 1. Insert the common pin (marked by a tab) on the dual banana connector into the negative (black) terminal marked with a "–" on the module. 2. Insert the opposite pin into the positive (red) terminal marked with a "+" on the side labeled "V" for voltage measurement and "mA" for milliamp measurement.
BetaPort–P Pressure Modules The BetaPort– P pressure modules are fitted with 316 S.S. 1/8" FNPT connections. A variety of connectors and adapters are available from third parties to allow connection to any type of field device. It is recommended that you use an appropriate pipe thread seal (such as Teflon tape) with any connector. This improves thread sealing and reduces the likelihood of galling the threads.
CAUTION: Failure to use a backup wrench may damage the calibrator and/or the module and will void the product warranty. Do not use the pressure connections to support the calibrator. Pressure Module Media Compatibility BetaPort–P pressure modules consist of two basic types: non-isolated isolated Consult the calibrator specifications (Appendix B) for details about which modules are isolated and which are non-isolated.
Pressure Module Safety Always use good safety practices when working with high pressures. Although liquids are considered noncompressible, dissolved and trapped gases can exist in the system and can store significant energy when compressed under high pressure. Always use good quality fittings that are rated for the pressures you are using and always use safety glasses or goggles. Overpressure Protection The sensors used in BetaPort–P pressure modules have excellent overpressure specifications.
Input Modules 2-8
3- Basic Operations Overview The BetaGauge II is a two-channel device. This means that two variables can be measured and displayed simultaneously. The display data for the channels 1 and 2 corresponds to the modules installed in the left and right module bays respectively. Channel 1 appears in the top half of the LCD display while Channel 2 appears in the bottom half. The BetaGauge II uses pressure modules to measure pressure and DC modules to measure electrical signals.
Selecting a Channel for Specific Functions Some functions, such as zeroing a module or setting a damping feature, require that you specify a channel prior to performing the function. NOTE: All sample displays representing the LCD in this guide are depicted with a 0 to 30 PSI gauge pressure module and/or a DC module. Procedure To select a channel for a specific function, press either the CH 1 or CH 2 key on the keypad. The selected channel appears in the lower right corner of the LCD display.
Selecting Pre-Programmed Display Units The BetaGauge II allows you to use pre-programmed display units for calibrating your field instrument. For each module installed, the BetaGauge II allows you to select the units used to display data on the LCD. The following paragraphs describe how to select display units for both types of modules (BetaPort–DC and BetaPort–P pressure). BetaPort–DC Module The BetaPort– DC module can measure in milliamps or Volts.
BetaPort–P Pressure Module The BetaPort– P pressure module can display in any of the following units: PSI InHg InWC mmHg cmWC kPa bar mbar atm kpcm MPa (User) Upon initialization, the BetaPort–P pressure module defaults to PSI pressure units. Use the following procedure to select other pre-programmed pressure units: 1. Press CH 1 or CH 2, as appropriate, to select the BetaPort–P pressure module for modification. 2.
NOTE: The last selection in the list of pressure units is User, which allows you to define your own pressure unit for display purposes. Refer to Section 5 for more information. 4. To return to the Setup menu, press the CLR key. Selecting a Water Column Measurement In the list of available display units for a BetaPort–P pressure module, InWC and cmWC are water column equivalent measurements.
Procedure To select a water column pressure unit, do the following: 1. Follow the procedure for selecting pre-programmed pressure units, selecting InWC or cmWC, as appropriate. The display screen appears similar to the one shown below: CH2 4C InWC 20C 60F 2. Press the corresponding function key to select the appropriate temperature for the water density correction value you want to use. The display returns to the list of available units. 3. Press the CLR key to return to the Setup menu.
Zeroing Pressure Modules The BetaGauge II has two different zeroing procedures for pressure modules, depending on the zero reference type installed. The two zero reference types include: Gauge zero. The gauge zero reference type modules have range labels with markings of PSIG, “H 20, or PSID. Absolute zero. The absolute zero reference modules have range labels with the marking of PSIA.
4. Ensure that the BetaGauge II is displaying the first four Setup menu items (Trip, Unit, Damp, and V/mA). NOTE: Only this menu display allows for both channel selection and ZERO key operation. 5. Press the channel select key corresponding to the module bay where the gauge zero reference module is installed (CH1 for the left side, or CH2 for the right side). 6. Press the ZERO key. 7. The pressure display should now be zeroed.
4. Ensure that the BetaGauge II is displaying the first four Setup menu items (Trip, Unit, Damp, and V/mA). NOTE: Only this menu display allows for both channel selection and ZERO key operation. 5. Press the channel select key corresponding to the module bay where the gauge zero reference module is installed (CH1 for the left side, or CH2 for the right side). 6. Press the ZERO key, and the lower portion of the LCD will prompt for Password. 7.
Zeroing DC Modules The BetaGauge II has two different zeroing procedures for DC modules, depending on the type of electrical signal to be measured: Volts or milliamps. Zeroing the Volts Input Use the following procedure to zero the Volts input on the DC module: 1. Install the module in either module bay (channel) of the BetaGauge II. 2. Ensure that the BetaGauge II is on. 3. Install the dual banana jack on the test leads into the positive Volts jack and the common jack. 4.
Zeroing the Milliamp Input Use the following procedure to zero the milliamp input on the DC module: 1. Install the module in either module bay (channel) of the BetaGauge II. 2. Ensure that the BetaGauge II is on. 3. Ensure that the module’s milliamp jacks are open circuit. This is done by ensuring that no test leads are attached to either the positive milliamp jack or the common jack. 4. Ensure that the BetaGauge II is displaying the first four Setup menu items (Trip, Unit, Damp, and V/mA).
Setting the Damping Factor for Ch1/Ch2 The BetaGauge II allows you to enable a damping feature to reduce the display fluctuations caused by a noisy input signal (pressure or electrical). NOTE: The BetaGauge II automatically measures (or reads) each module three times per second. The following settings are available: Damping Setting The BetaGauge II … 0.0 (Disabled) Updates the displayed value with every measurement (three times per second). 2.
Reset Damping Guidelines The following reset damping guidelines apply: The BetaGauge II automatically disables (resets to 0.0 seconds) the damping feature for both channels each time you turn the calibrator off and on. The BetaGauge II automatically disables (resets to 0.0 seconds) the damping feature for a particular channel each time you install, or re-install, a module into the module bay corresponding to that channel.
3. Press the function key corresponding to the damping setting you want to use, then press the CLR key to return to the Setup menu. CAUTION: Enabling the damping feature affects your ability to accurately zero a module’s displayed value. Therefore, prior to zeroing the displayed value of a module, disable the damping feature by setting the damping value to 0.0.
Pressure Module Example: GAGE SN: 0000000003 FS= 15.00 PSI Cal: 04/16/2005 3. Press the CLR key to return to the Module Info menu. 4. Press the CLR key again to return to the Setup menu. Programming Trip Detection The BetaGauge II has the ability to detect trips from either wet or dry contacts using a BetaPort–DC module. However, this trip detect feature is intended for use with pressure trip devices only. CAUTION: Do not install two BetaPort–DC electrical modules in the calibrator simultaneously.
Dry Contacts To detect dry contact trips, you must connect dry contacts to the milliamp input of the calibrator’s DC module. The BetaGauge II will detect a dry contact trip when the input resistance crosses the 800 Ohm threshold.
Pressure Source Adjustment Perform the following tasks to adjust the pressure source prior to starting a pressure trip test: 1. Learn the direction (low-going or high-going) in which the trip will occur and approximately where the device is expected to trip (in terms of pressure). 2. Use the coarse adjustment control of the pressure source to set the pressure 5 to10% away from the trip point. 3. Use the fine adjustment control of the pressure source to ramp through the expected pressure trip point.
Preliminary Setup Use the following procedure to set up the BetaGauge II for trip detection: 1. Connect the alligator clip ends of the test leads (supplied with the BetaGauge II) to the contacts you want to monitor. 2. Do one of the following: If the contacts are wet (have voltage on one or both sides), connect the dual banana jack to the voltage input of the DC module.
Procedure Use the following procedure to program a trip detection: 1. Press CH 1 or CH 2, as appropriate, to activate the BetaPort–DC module. 2. Use the Menu key to locate the Setup menu item Trip, then press the corresponding function key to display a screen similar to the one shown below: CHx Trip Disabled Dis. Enab Reset In the above example: x = the channel number of the DC module. 3.
7. If possible, use the micro-adjustment of the pressure source to ramp towards the reset point of the pressure trip device. The display will again display TRIP when the BetaGauge II detects the reset point. 8. When you have completed trip detect testing, do the following: Vent the pressure source. Disconnect all test circuits. Press the function key corresponding to Dis. to disable the trip detection feature. Press the CLR key to return to the Setup menu.
4- User Preferences Overview There are many BetaGauge II preferences settings that you can program to meet your special needs. Although you can change these settings as often as you like, most users set them once when they acquire their calibrator.
Programming and Viewing Calibrator Information The BetaGauge II allows you to enter the owner’s name (or other identification) into the calibrator’s nonvolatile memory. You can then view this identification, as well as the following information, upon demand: Serial number of the calibrator Calibrator’s last factory acceptance date Entering Owner Identification To program the owner’s name (or other identification) into the calibrator’s non-volatile memory, do the following: 1.
Viewing Calibrator Information To view calibrator information, do the following: 1. Use the Menu key to locate the Setup menu item Owner, then press the corresponding function key. The following data appears in the lower portion of the LCD: Owner’s name Serial number of the calibrator Calibrator’s last factory acceptance date A sample is shown below: MARTEL SN:17009 Cal. Date: 04/25/2006 2. To return to the Setup menu, press the CLR key.
2. To disable the beep sound, press the function key corresponding to the menu item Dis. 3. To enable the beep sound, press the function key corresponding to the menu item Enab. 4. To clear the display and return to the Setup menu, press the CLR key. Setting LCD Preferences The BetaGauge II allows you to set the following LCD variables: Specify the length of time the LCD backlight will remain illuminated Set the LCD contrast Perform a contrast recovery procedure to restore the display.
2. Press the function key corresponding to the menu item LED. The lower portion of the LCD will appear similar to the example shown below: Set backlit time 33 3. seconds Use the numbers on the keypad to enter a backlight display value between 0 and 3600 seconds, then press the ENTER key to return to the LCD menu. NOTE: If you attempt to enter a value outside this range, an Invalid Entry message will appear.
3. Press the menu items Light or Dark as many times as necessary to adjust the LCD contrast. 4. When you have completed setting the contrast, press the CLR key to return to the LCD menu. 5. Press the CLR key again to return to the Setup menu. Recovering the LCD Contrast Extreme hot and/or cold temperatures can affect the LCD contrast.
Setting the Automatic Shut Off Delay Time To prevent the battery pack from discharging while the calibrator is not in use, you can program the BetaGauge II to automatically shut off after a pre-defined period of idle time. Idle time is defined as a period with no key presses. You can also disable this feature if necessary. Procedure To set the automatic shut off delay time, do the following: 1. Use the Menu key to locate the Setup menu item Off, then press the corresponding function key.
5. Use the numbers on the keypad to enter a delay value between 200 and 3600 seconds, then press the ENTER key to return to the Auto Off menu. NOTE: If you attempt to enter a value outside this range, an Invalid Entry message will appear. To erase an existing entry or to correct a mistake, use the CLR key to erase the character to the left of the cursor. 6. From the Auto Off menu, press the CLR key to return to the Setup menu.
2. Press the function key corresponding to the menu item Sec to display the Set Second screen as shown below: Set Second 3. Use the number keys on the keypad to enter a value between 0 and 59 seconds, then press the ENTER key to return to the Set time menu. NOTE: If you attempt to enter a value outside this range, an Invalid Entry message will appear. To erase an existing entry or to correct a mistake, use the CLR key to erase the character to the left of the cursor. 4.
2. The rightmost menu item allows you to toggle between 12 and 24-hour format. Press the function key corresponding to the menu item 12Hr or 24Hr, as appropriate. NOTE: Notice how the display time in the upper left portion of the LCD changes between 12 and 24 hour format as you toggle between the two. 3. Press the CLR key to return to the Setup menu. Setting and Formatting the Date Display The BetaGauge II allows you to manually set the date as well as format how the date will display.
2. Press the function key corresponding to the menu item Day to display the Set Day screen as shown below: Set Day 3. Use the numbers on the keypad to enter the day of the month between 0 and 31, then press the ENTER key to return to the Set Date menu. NOTE: If you attempt to enter a value outside this range, an Invalid Entry message will appear. To erase an existing entry or to correct a mistake, use the CLR key to erase the character to the left of the cursor. 4.
Procedure To specify the date format, do the following: 1. Use the Menu key to locate the Setup menu item Format, then press the corresponding function key. The lower portion of the LCD will appear as shown below: Format Date time dmy mdy ymd 12Hr 2. The first three menu items allow you specify the date format you want to use. Press the function key corresponding to the menu item dmy, mdy, or ymd, as appropriate.
5- Advanced Display Modes Overview The BetaGauge II allows you to define advanced display modes for both the BetaPort–P pressure and BetaPort–DC modules. This section describes how to set up these advanced display modes. NOTE: For information on how to use preprogrammed display units, refer to Section 3. Programming a User Defined Pressure Unit If none of the eleven (11) pre-programmed pressure units (listed in Section 3) meet your needs, the BetaGauge II allows you to program your own pressure unit.
Procedure Use the following procedure to define a pressure unit of your choice: 1. Press CH1 or CH2 to indicate the appropriate bay in which the pressure module is installed. 2. Use the MENU key to locate the Setup menu item User, then press the corresponding function key to display the following screen: CHx Ent. Unit: yyyy In the above example: x = the channel number of the pressure module, yyyy = the current user unit label. 3.
4. When the unit label is satisfactory, press the ENTER key once again to display the following screen: Ent. 1PSI = 5. Use the numbers on the keypad to specify the 1 PSI equivalent value of your user unit, then press the ENTER key to return to the Setup menu. NOTE: For the example Feet of Water Column at 20°C, you would enter 2.3108. 6. Use the MENU key to locate the Setup menu item Unit, then press the corresponding function key to display a screen similar to the following: CHx mmHg cmWC Kpa → 7.
Programming Display Modes for BetaPort–DC Modules The BetaGauge II allows you to define process signals for display using any of the following modes: engineering units percent scale percent error square root extraction NOTE: Before performing any of the procedures for setting up display modes for BetaPort–DC modules, ensure that you are familiar with the input and output signal ranges of the field instrument you are calibrating.
Defining a Unit Label To define a unit label, do the following: 1. Ensure that a BetaPort–DC module is installed in one of the module bays, then press CH1 or CH2 to indicate the appropriate bay in which the DC module is installed. 2. If necessary, use the V/mA Setup menu item to change the electrical units to match the field instrument's electrical signal. 3. Use the Menu key to locate the Setup menu item User, then press the corresponding function key to display the following screen: CHx Ent.
Defining the Field Instrument Type To define the type of field instrument you want to calibrate, do the following: 1. Use the Menu key to locate the Setup menu item Mode, then press the corresponding function key to display the following screen: Calib Instr P/I I/P Other 2. Press the corresponding function key to select the type of field instrument you want to calibrate: P/I, I/P, or Other. 3. If you select Other, the following screen will display: Instrument Input: CH1 4.
Defining Engineering Low/High Values To define the engineering low and high values with respect to the expected field instrument output signals, do the following: 1. After you define the field instrument type, the following main linearization screen displays: CHx Mode: yyyy Std Eng → Sqrt In the above example: x = the channel number of the DC module, yyyy = the current linearization mode of the DC module channel. 2.
5. After you have specified all of the values in step 4, the display screen prompts you for the number of decimal places you want to display. A sample screen is shown below: CHx Ent. Decimals: 6. Use the keypad numbers to enter the number of decimal places you want to appear on the display (from 1 to 3), then press the ENTER key to return to the Calib Instr menu. 7. Press the CLR key to return to the Setup menu.
Defining the Field Instrument Type To define the type of field instrument you want to calibrate, do the following: 1. Ensure that a BetaPort–DC module is installed in one of the module bays, then press CH1 or CH2 to indicate the appropriate bay in which the DC module is installed. 2. Use the Menu key to locate the Setup menu item %Fnc, then press the corresponding function key to display the following screen: Calib Instr P/I I/P Other 3.
6. After you select the field instrument type, the main linearization screen displays. Use this screen to begin defining the expected low and high electrical signal values of the field instrument as described in the next procedure. Defining the Expected Low/High Values To define the expected low and high electrical signal values of the field instrument, do the following: 1.
4. Use the keypad numbers to enter a value equal to the expected low electrical signal value of the field instrument, then press the ENTER key. The display screen then prompts you for an expected high electrical signal value. 5. Repeat step 4 to enter a value for the expected high electrical signal value. 6. After you have specified values for both the expected low and high electrical signals, the display screen will return to the main linearization screen.
Defining the Field Instrument Type To define the type of field instrument you want to calibrate, do the following: 1. Ensure that a BetaPort–DC module is installed in one of the module bays, then press CH1 or CH2 to indicate the appropriate bay in which the DC module is installed. 2. Use the Menu key to locate the Setup menu item %Fnc, then press the corresponding function key to display the following screen: Calib Instr P/I I/P Other 3.
6. After you select the field instrument type, the main linearization screen displays. Use this screen to begin defining the expected low and high values for the field instrument’s input and output scales as described in the next procedure. Defining the Expected Low/High Values 1.
5. The display screen individually prompts you for the following values: expected signal value for the high end of the field instrument’s output scale expected signal value for the low end of the field instrument’s input scale expected signal value for the high end of the field instrument’s input scale Repeat step 4 to enter values for each of the above prompts. 6. After you have specified all of the values in step 5, the display screen will return to the main linearization screen.
Square Root Extraction The BetaGauge II is capable of extracting square roots from flow transmitters or other devices that output a squared output signal. Before setting up a square root extraction display, you must perform the following procedures: Define a unit label, or name. Define the expected output milliamp signal for flows that are equal to 0% and 100%, and define both the low and high end flow variables.
3. Specify a unit label by using the CLR key to erase the character (if any) to the left of the cursor and use the arrow keys on the keypad to scroll through the available character set. When the desired character is displayed, press the ENTER key to accept the character and move the cursor to the right. A maximum of 4 characters may be entered for identification. NOTE: Pressing the ENTER key twice in a row ends editing.
3. Press the function key corresponding to Sqrt and the display screen prompts you for the field instrument’s expected output signal for a 0% flow. A sample screen is shown below: CHx Instr. Out Lo 4. Use the keypad numbers to enter the value of the expected output milliamp signal when the flow is equal to 0% (typically 4 for a 4-20 milliamp output). Press the ENTER key. 5.
7. Use the numbers on the keypad to enter the number of decimal places (1 to 3 ) you want to appear in the display, then press the ENTER key to return to the Calib Instr menu. 8. Press the CLR key to return to the Setup menu.
6- Application Notes Overview The BetaGauge II allows you to perform many different calibration tests. In addition, the configurations of these tests may vary, depending on whether you are using the conventional calibrator mode or documenting calibrator (SYS) mode of operation. Modes of Operation The BetaGauge II has two modes of operation: the conventional calibrator mode and the documenting calibrator (SYS) mode.
Other Necessary Equipment When using the BetaGauge II to calibrate your field instruments, it is necessary that you have certain equipment at your disposal. This equipment must be capable and accurate to result in an accurate calibration.
Calibration Tests You can perform many different calibration tests using the BetaGauge II.
Electrical to Pressure Test You can perform an electrical to pressure calibration test using either the conventional mode or the SYS mode of operation. The following paragraphs describe all of the possible configurations, as well as, the other necessary equipment to perform an electrical to pressure test.
Illustration A graphical representation for configuring an electrical to pressure calibration test using the conventional mode of operation is shown below: Figure 6-1 Electrical to Pressure Calibration Test Using Conventional Mode 6-5 Application Notes
SYS Mode There are two methods for configuring an electrical to pressure calibration test using SYS mode. The following table indicates which method to use. Use this method if… Method 1 Your field instrument has a pressure output range within the ability of the pressure module on the BetaGauge II. Method 2 Your field instrument has a pressure output range that exceeds the ability of the pressure module on the BetaGauge II.
Illustration A graphical representation for configuring the Method 1 electrical to pressure calibration test using the SYS mode of operation is shown below: Figure 6-2 Electrical to Pressure Calibration Test Using SYS Mode (Method 1) 6-7 Application Notes
Method 2 Software Instrument Setup Electrical input (requires manual entry on the BetaGauge II) Pressure output (requires manual entry on the BetaGauge II) Other Necessary Equipment Electrical source Pressure measurement device Application Notes 6-8
Illustration A graphical representation for configuring the Method 2 electrical to pressure calibration test using the SYS mode of operation is shown below: Figure 6-3 Electrical to Pressure Calibration Test Using SYS Mode (Method 2) 6-9 Application Notes
Pressure to Electrical Test You can perform a pressure to electrical calibration test using either the conventional mode or the SYS mode of operation. The following paragraphs describe all of the possible configurations, as well as, the other equipment necessary to perform a pressure to electrical test.
Illustration A graphical representation for configuring a pressure to electrical calibration test using the conventional mode of operation is shown below: Figure 6-4 Pressure to Electrical Calibration Test Using Conventional Mode 6-11 Application Notes
SYS Mode There are four methods for configuring a pressure to electrical calibration test using SYS mode. The following table indicates which method to use. Use this method if… Method 1 Your field instrument has a pressure output range within the ability of the pressure module on the BetaGauge II, and Your field instrument has an electrical output range within the ability of the DC module on the BetaGauge II.
Use this method if… Method 4 Your field instrument has a pressure input range that exceeds the ability of the pressure module on the BetaGauge II, and Your field instrument has an electrical output range that exceeds the ability of the DC module on the BetaGauge II.
Method 1 Software Instrument Setup Pressure input Electrical output Other Necessary Equipment Pressure source Application Notes 6-14
Illustration A graphical representation for configuring the Method 1 pressure to electrical calibration test using the SYS mode of operation is shown below: Figure 6-5 Pressure to Electrical Calibration Test Using SYS Mode (Method 1) 6-15 Application Notes
Method 2 Software Instrument Setup Pressure input (requires manual entry on the BetaGauge II) Electrical output Other Necessary Equipment Pressure source Pressure measurement device Application Notes 6-16
Illustration A graphical representation for configuring the Method 2 pressure to electrical calibration test using the SYS mode of operation is shown below: Figure 6-6 Pressure to Electrical Calibration Test Using SYS Mode (Method 2) 6-17 Application Notes
Method 3 Software Instrument Setup Pressure input Electrical output (requires manual entry on the BetaGauge II) Other Necessary Equipment Pressure source Electrical measurement device Application Notes 6-18
Illustration A graphical representation for configuring the Method 3 pressure to electrical calibration test using the SYS mode of operation is shown below: Figure 6-7 Pressure to Electrical Calibration Test Using SYS Mode (Method 3) 6-19 Application Notes
Method 4 Software Instrument Setup Pressure input (requires manual entry on the BetaGauge II) Electrical output (requires manual entry on the BetaGauge II) Other Necessary Equipment Pressure source Pressure measurement device Electrical measurement device Application Notes 6-20
Illustration A graphical representation for configuring the Method 4 pressure to electrical calibration test using the SYS mode of operation is shown below: Figure 6-8 Pressure to Electrical Calibration Test Using SYS Mode (Method 4) 6-21 Application Notes
Electrical to Electrical Test You can perform an electrical to electrical calibration test using only the SYS mode of operation. The following paragraphs describe the possible configurations, as well as, the other equipment necessary to perform an electrical to electrical test.
SYS Mode There are two methods for configuring an electrical to electrical calibration test using SYS mode. The following table indicates which method to use. Use this method if… Method 1 Method 2 Your field instrument has an electrical output range within the ability of the DC module on the BetaGauge II. Your field instrument has an electrical output range that exceeds the ability of the DC module on the BetaGauge II.
Method 1 Software Instrument Setup Electrical input (requires manual entry on the BetaGauge II) Electrical output Other Necessary Equipment Electrical source Application Notes 6-24
Illustration A graphical representation for configuring the Method 1 electrical to electrical calibration test using the SYS mode of operation is shown below: Figure 6-9 Electrical to Electrical Calibration Test Using SYS Mode (Method 1) 6-25 Application Notes
Method 2 Software Instrument Setup Electrical input (requires manual entry on the BetaGauge II) Electrical output (requires manual entry on the BetaGauge II) Other Necessary Equipment Electrical source Electrical measurement device Application Notes 6-26
Illustration A graphical representation for configuring the Method 2 electrical to electrical calibration test using the SYS mode of operation is shown below: Figure 6-10 Electrical to Electrical Calibration Test Using SYS Mode (Method 2) 6-27 Application Notes
Pressure to Pressure Test You can perform a pressure to pressure calibration test using either the conventional mode or the SYS mode of operation. The following paragraphs describe all of the possible configurations, as well as, the other equipment necessary to perform a pressure to pressure test.
Illustration A graphical representation for configuring a pressure to pressure calibration test using the conventional mode of operation is shown below: Figure 6-11 Pressure to Pressure Calibration Test Using Conventional Mode 6-29 Application Notes
SYS Mode There are three methods for configuring a pressure to pressure calibration test using SYS mode. The table on the following page indicates which method to use.
Use this method if… Method 1 Your field instrument has a pressure input range within the ability of one pressure module on the BetaGauge II, and Your field instrument has a pressure output range within the ability of the other pressure module on the BetaGauge II.
Method 1 Software Instrument Setup Pressure input Pressure output Other Necessary Equipment Pressure source Application Notes 6-32
Illustration A graphical representation for configuring the Method 1 pressure to pressure calibration test using the SYS mode of operation is shown below: Figure 6-12 Pressure to Pressure Calibration Test Using SYS Mode (Method 1) 6-33 Application Notes
Method 2 Software Instrument Setup Pressure input Pressure output (requires manual entry on the BetaGauge II) Other Necessary Equipment Pressure source Pressure measurement device Application Notes 6-34
Illustration A graphical representation for configuring the Method 2 pressure to pressure calibration test using the SYS mode of operation is shown below: Figure 6-13 Pressure to Pressure Calibration Test Using SYS Mode (Method 2) 6-35 Application Notes
Method 3 Software Instrument Setup Pressure input (requires manual entry on the BetaGauge II) Pressure output Other Necessary Equipment Pressure source Pressure measurement device Application Notes 6-36
Illustration A graphical representation for configuring the Method 3 pressure to pressure calibration test using the SYS mode of operation is shown below: Figure 6-14 Pressure to Pressure Calibration Test Using SYS Mode (Method 3) 6-37 Application Notes
Temperature to Electrical Test You can perform a temperature to electrical calibration test using only the SYS mode of operation. The following paragraphs describe the possible configurations, as well as, the other equipment necessary to perform a temperature to electrical test.
SYS Mode There are two methods for configuring a temperature to electrical calibration test using SYS mode. The following table indicates which method to use. Use this method if… Method 1 Method 2 Your field instrument has an electrical output range within the ability of the DC module on the BetaGauge II. Your field instrument has an electrical output range that exceeds the ability of the DC module on the BetaGauge II. NOTE: The BetaGauge II cannot make direct temperature measurements.
Method 1 Software Instrument Setup Temperature input (requires manual entry on the BetaGauge II) Electrical output Other Necessary Equipment Temperature source Application Notes 6-40
Illustration A graphical representation for configuring the Method 1 temperature to electrical calibration test using the SYS mode of operation is shown below: Figure 6-15 Temperature to Electrical Calibration Test Using SYS Mode (Method 1) 6-41 Application Notes
Method 2 Software Instrument Setup Temperature input (requires manual entry on the BetaGauge II) Electrical output (requires manual entry on the BetaGauge II) Other Necessary Equipment Temperature source Electrical measurement device Application Notes 6-42
Illustration A graphical representation for configuring the Method 2 temperature to electrical calibration test using the SYS mode of operation is shown below: Figure 6-16 Temperature to Electrical Calibration Test Using SYS Mode (Method 2) 6-43 Application Notes
Electrical Indicator or Loop Test You can perform an electrical indicator or loop calibration using either the conventional mode or the SYS mode of operation. The following paragraphs describe all of the possible configurations, as well as, the other equipment necessary to perform an electrical indicator test. Conventional Mode You may perform this test if: Your field instrument has an electrical input range within the ability of the DC module on the BetaGauge II.
Illustration A graphical representation for configuring an electrical indicator or loop calibration using the conventional mode of operation is shown below: Figure 6-17 Electrical Indicator or Loop Calibration Using Conventional Mode 6-45 Application Notes
SYS Mode You may perform this test if: You are using either a local or remote electrical indicator to indicate the electrical input signal to the field instrument. NOTE: While in SYS mode, the BetaGauge II cannot measure the electrical input signals to the field instrument. You must manually enter the value of these signals on the BetaGauge II.
Illustration A graphical representation for configuring the electrical indicator or loop calibration using the SYS mode of operation is shown below: Figure 6-18 Electrical Indicator or Loop Calibration Using SYS Mode 6-47 Application Notes
Pressure Indicator or Loop Test You can perform a pressure indicator or loop calibration using either the conventional mode or the SYS mode of operation. The following paragraphs describe all of the possible configurations, as well as, the other equipment necessary to perform a pressure indicator test.
Illustration A graphical representation for configuring pressure indicator or loop calibration using the conventional mode of operation is shown below: Figure 6-19 Pressure Indicator or Loop Calibration Using Conventional Mode 6-49 Application Notes
SYS Mode There are two methods for configuring a pressure indicator or loop calibration using SYS mode. The following table indicates which method to use. Use this method if… Method 1 Your field instrument has a pressure input range within the ability of the pressure module on the BetaGauge II, and You are using either a local or remote pressure indicator to indicate the pressure input signal to the field instrument.
Illustration A graphical representation for configuring the Method 1 pressure indicator or loop calibration using the SYS mode of operation is shown below: Figure 6-20 Pressure Indicator or Loop Calibration Using SYS Mode (Method 1) 6-51 Application Notes
Method 2 Software Instrument Setup Pressure input (requires manual entry on the BetaGauge II) Pressure indicator output (requires manual entry on the BetaGauge II) Other Necessary Equipment Pressure source Pressure measurement device Application Notes 6-52
Illustration A graphical representation for configuring the Method 2 pressure indicator or loop calibration using the SYS mode of operation is shown below: Figure 6-21 Pressure Indicator or Loop Calibration Using SYS Mode (Method 2) 6-53 Application Notes
Temperature Indicator or Loop Test You can perform a temperature indicator or loop calibration using only the SYS mode of operation. The following paragraphs describe the possible configurations, as well as, the other equipment necessary to perform a temperature indicator test. SYS Mode You may perform this test if: You are using either a local or remote temperature indicator to indicate the temperature input signal to the field instrument.
Illustration A graphical representation for configuring the temperature indicator or loop calibration using the SYS mode of operation is shown below: Figure 6-22 Temperature Indicator or Loop Calibration Using SYS Mode 6-55 Application Notes
Application Notes 6-56
7- Documenting Calibrator (SYS) Mode Overview Typical instrument maintenance software packages allow you to track and maintain your facility’s instruments and their calibration histories. Using the BetaGauge II SYS mode and an instrument maintenance software package, the calibrator automatically prompts you for pre-defined test points, then documents the instrument's output and input signals for later upload to a personal computer.
Before You Begin Before you begin using the SYS mode to calibrate your field instrument(s), there are two tasks you must perform: Download Instrument Tests Connect Necessary Equipment Downloading Instrument Tests Before you can access the BetaGauge II SYS mode, you must first download instrument tests from a software database on a personal computer to the BetaGauge II. These instrument tests are referred to as tags.
Connecting Necessary Equipment Make sure that you have connected all of the necessary equipment required for the calibration test you want to perform. This equipment includes the field instrument you want to calibrate, the BetaGauge II, as well as, any other equipment necessary to performing a particular calibration test. For information on how to configure the equipment, refer to Section 6, Application Notes.
Back Screen Level You can access the Back Screen level via the Main level only. From the Back Screen level, you can perform the following functions: Access the Execute level (described below). Erase all the data recorded for a selected tag. NOTE: The erase all data function is available only if calibration reset is allowed in your instrument maintenance software. Return to the Main level. Execute Level You can access the Execute level via the Back Screen level only.
Accessing the Main Level To access the Main level of the SYS mode, use the Menu key to locate the Setup menu item Sys, then press the corresponding function key to display the Tag Status screen showing the first available tag. A sample display is shown below: Tag 1: BG2 TAGS Tag 1 Inst. OUT: mA Inst. IN: PSI Status: UNCAL Time: 00:00:00 Date: 00/00/00 Error Tol: 0.
you must return to the Tag Status screen to specify a different tag. The Tag Status screen is divided into six sections. These sections are described below in the order that they appear on the screen from top to bottom: Tag Number/Area. The current tag number and area is listed in the top portion of the screen. NOTE: When using any instrument maintenance software packages, all tags may be assigned to an area. Refer to your instrument maintenance software manual for more information. Input/Output Types.
Menu Items The menu items shown on the bottom of the Tag Status screen are described below: Menu Item Allows you to … Tag Specify a particular tag for viewing or calibration purposes. TP View the expected input and output measurements of each test point for the current tag. Rslt View the calibration or test results of the last test performed on the current tag.
Viewing the Test Points for a Specific Tag To view the test points for a specific tag on the Tag Status screen, do the following: 1. Specify the tag number associated with the test points you want to view. 2. Press the function key corresponding to the menu item TP to display the expected input and output measurement for the first six test points of the current tag. An example screen is shown below: Tag 2: BG2 TAGS Tag 2 Inst. OUT: mA Inst. IN: PSI Pt. Exp. IN Exp. OUT Un 1 3.0000 4.0000 Un 2 6.0000 8.
Viewing the Calibration Results for a Specific Tag When you have completed calibrating a specific tag, you can view the calibration results on the BetaGauge II. NOTE: The BetaGauge II will display the calibration results from the last calibration performed on the current tag. For information on how to calibrate test points, see the procedure titled Tag Calibration later in this section. From the Tag Status screen use the following procedure to view the last calibration results for a specific tag: 1.
4. To return to the Tag Status screen, press the CLR key. NOTE: You can exit the SYS mode by pressing the CLR key again. Accessing the Back Screen Level To access the Back Screen level of the SYS mode, press the function key corresponding to → on the Tag Status screen. A sample Back screen display is shown below: Tag 1: BG2 TAGS Tag 1 Inst. OUT: mA Inst. IN: PSI Status: UNCAL Time: 00:00:00 Date: 00/00/00 Error Tol: 0.
Menu Items The menu items shown on the bottom of the Back screen are described below: Menu Item Allows you to … Exe Access the Execute level (described on the following pages). Reset Erase all recorded data for the selected tag. NOTE: The erase all data function is available only if calibration reset is allowed in your instrument maintenance software. → Return to the Main level. 7-11 Doc. Cal.
Erasing Data for a Specific Tag To erase all recorded data for a specific tag on the Back screen, do the following: NOTE: Recorded data exists only if you have performed calibration tests for the selected tag. For information on how to perform a calibration test, see the procedure titled Tag Calibration later in this section. 1. Specify the tag number for the data you want to erase. 2.
Accessing the Execute Level To access the Execute level of the SYS mode, press the function key corresponding to Exe on the Back screen. The Test Setup screen appears as shown below: Tag 2: BG2 TAGS Tag 2 Inst. OUT: mA 4.0000 20.000 CH1 Inst. IN: PSI 3.0000 15.000 CH2 CH1 DC 09/25/1995 CH2 GAGE 04/16/1995 15.00 Modules Correct Ready The Test Setup screen is described below and on the following page.
The Test Setup screen is divided into five sections. These sections are described below in the order that they appear on the screen from top to bottom: Tag Number/Area. The current tag number and area is listed in the top portion of the screen. Low/High Test Point Values. This section lists the lowest and highest test point values of the selected tag. These values include the minimum to maximum ranges for the field instrument’s input and output measurements.
Choosing Appropriate Modules In choosing the appropriate modules for the test point values of the current tag, you can use the following examples as guidelines. Example 1: Over-Pressure Protection For those tags that require a pressure module, the BetaGauge II will accept a 5000 PSI module to perform a 5 PSI test. However, if you install a 5 PSI module for a test that requires 200 inches of water column, the status message will read Wrong Module(s).
Accessing the Zero Screen When the Modules Correct status message appears on the Test Setup screen, you can access the Zero screen where you can zero the display readings of the installed modules. To access the Zero screen, press the function key corresponding to the Ready menu item on the Test Setup screen. The Zero screen shows live input displays of the two modules currently installed. A sample is shown below: Tag 2: BG2 TAGS Tag 2 CH1 0.0008 Inst. OUT mA DC CH2 0.1103 Inst.
System Prompt. This section first prompts you to zero the display for channel 1, then follows with channel 2. Menu Items. The menu items (Ready and Zero) that appear at the bottom of the screen allow you to zero the display for the current channel or bypass the zeroing operation for the current channel. Zeroing the Display Readings From the Zero screen, use the following procedure to zero the display readings of the installed modules prior to starting a test.
2. After you select the Zero or Ready menu item for channel 1, the calibrator prompts you to zero channel 2. Repeat step 1 to zero the display for channel 2 or bypass the option. 3. After you select the Zero or Ready menu item in step 2, the calibrator displays the Execute screen. Refer to the paragraphs below for information concerning the Execute screen. NOTE: You can exit the SYS mode by pressing the CLR key once to return to the Tag Status screen, then once again to return to the Setup menu.
Execute Screen Description The Execute screen is divided into six sections. These sections are described below in the order that they appear on the screen from top to bottom: Tag Number/Area. The current tag number and area is listed in the top portion of the screen. NOTE: When using an instrument maintenance software package, all tags must be assigned to an area. Refer to your instrument maintenance software manual for more information. Input/Output Types.
Tag Calibration Before you begin the calibration process, it is important to know the following information: You must have already specified the tag number you want to calibrate on the Tag Status screen in the Main level. (In other words, you cannot specify a tag number from the Execute screen.) You can cancel the calibration test at any time during the process by pressing the CLR key to return to the Tag Status screen. To perform tag calibration, do the following: 1.
3. When the applied field instrument input value comes within the allowed pressure tolerance, the BetaGauge II beeps. When you hear the beep, press the ENTER key. (The system will not accept the ENTER key until you hear the beep.) NOTE: If the actual field instrument output is within the allowed error tolerance of the calculated ideal output, the BetaGauge II will change the calibration status for that test point from Un to C.
Accessing the Post-Test Status Screen After you have calibrated the last test point for the current tag, the BetaGauge II displays the Post-Test Status screen. A sample is shown below: Tag 2: BG2 TAGS Tag 2 Inst. OUT: mA Inst. IN: PSI Status: PASS Time: 11:23:59 Date: 10/07/96 Error Tol: 0.0500 Alert Tol: 999999 Next: As Left 1 Adj. Instr.
Specifying Post-Test Status Information NOTE: You must answer the post-test status prompts. If you press the CLR key prior to answering the post-test prompts, all calibration results will be lost. To specify post-test status information, do the following; 1. From the Post-Test Status screen, use the function button corresponding to the appropriate menu item (Yes, No, or Rep) to indicate whether the field instrument was adjusted, left alone, or replaced. 2.
Doc. Cal.
A-Battery Care/Maintenance Overview Before handling the batteries used with the BetaGauge II, read the following danger, warning, and caution statements. DANGER! Failure to carefully observe the following procedures and precautions can result in battery leakage, heat generation, bursting, and serious personal injury. Never dispose of a battery pack in a fire or expose a battery pack to high temperatures.
WARNING! Do not apply water, seawater, or other oxidizing agents to the battery pack, as this can cause rust and heat generation and can result in the battery pack bursting. Never use the battery pack if it is leaking, deformed, discolored, damaged, or is different from its original condition. External damage to the battery pack can be a sign of a malfunction. When the usage time for the battery pack becomes extremely short after charging, its operating life has ended and it should be replaced.
CAUTION! Do not strike or drop the battery pack. Sharp impacts or concussions to the battery pack may result in leakage, heat generation, and bursting. Always follow the specified charging temperature ranges on the battery pack.
Battery Packs The BetaGauge II uses rechargeable nickel–cadmium (NiCd) batteries that have high power-to-size-andweight ratios. If properly maintained and cared for, they will provide years of excellent service. However, NiCd batteries are susceptible to damage under the following conditions: Heat. Protect the batteries from prolonged exposure to extremely hot temperatures. Cold. Protect the batteries from prolonged exposure to extremely cold temperatures.
Guidelines for Using Batteries It is recommended that you use the following guidelines when using batteries: When installing or removing a battery pack, do not touch or bend the battery connection terminals on the calibrator. When installing or removing a battery pack, do not short the terminals on the back of the battery. Doing so will damage the battery pack and may cause bodily injury. Dispose of unusable batteries as specified by local regulations.
Removing a Battery Pack Use the following procedure to remove a battery pack: 1. Before removing a battery pack, turn off the calibrator and disconnect the charger. 2. Loosen the two captive screws on the back of the battery pack. 3. Pull the top of the battery pack away from the calibrator, lifting up and out.
Viewing the Battery Voltage There are two methods for viewing the current battery voltage: Approximate reading. A battery symbol, in the upper right corner of the LCD display, shows the approximate battery voltage. The highlighted (or dark) portion of the battery indicates the amount of available voltage.
NOTE: The scale is 6.5 Volts when full and 5.5 Volts when empty. Nominal full charge is 7.00 Volts, and the calibrator shuts off at 5.3 Volts. The message Low Battery flashes when the battery voltage drops below 5.5 Volts. Exact reading. You can view the exact voltage of the current battery by pressing the BAT key on the keypad. When you press the BAT key, the battery voltage reading replaces the battery symbol in the upper right portion of the LCD.
Quick Charge Pack (optional) The quick charge battery pack is designed with special discharge circuitry that can cycle the battery to a discharged state before recharging. This cycling prevents battery failure due to the memory effect of NiCd batteries. The quick charge battery pack requires the high capacity quick charger for proper charging. The charger for the quick charge battery pack is a high capacity charger that can fully charge the battery pack from a fully discharged state within 90 minutes.
The status of the quick charge battery pack charge cycle is indicated by the LED indicator on the battery pack: Steady On indicates fast charging. Rapid Flash indicates charged. Slow Flash indicates discharging. Off indicates that the cycle is disabled or the battery pack is not receiving power from the AC adapter. NOTE: The charger for the quick charge battery pack is different from the standard charger. The two should not be used interchangeably.
B-Specifications Available Pressure Ranges All ranges listed below are for a specific input module. Additional modules are available as options. Custom ranges outside the minimum and maximum ranges listed are also available as options. Non-isolated Gauge (PSIG) 0 to 0.3 (20 mBar) 0 to 1 (70 mBar) Isolated Gauge (PSIG) 0 to 5 (0 to 350 mBar) † 0 to 7.2 (0 to 500 mBar) (200” WC)†† 0 to 10 (0 to 700 mBar) †† 0 to 30 (0 to 2 Bar) 0 to 50 (0 to 3.
Absolute (PSIA) 0 to 15 (0 to 1 Bar) †† 0 to 30 (0 to 2 Bar) 0 to 50 (0 to 3.5 Bar) 0 to 100 (0 to 7 Bar) Compound (PSIG) -15 to 15 (-1 to 1 Bar) †† -15 to -30 (-1 to 2 Bar) Differential (PSID) (Non-isolated) ** 0 to 5 (0 to 350 mBar) † 0 to 30 (0 to 2 Bar) 0 to 50 (0 to 3.5 Bar) ** Maximum static pressure, 200 psig (14 Bar) † Thermal and Pressure Hysteresis: 0.0030 PSI (2.068 mBar) †† Thermal and Pressure Hysteresis: 0.0025 PSI (0.
Other Pressure Units (Keypad Selectable) kp/cm 2 (kilograms per sq. cm) Atmos (atmospheres) mbar (millibars) bar kPa (kiloPascals) MPa (MegaPascals) InHg (inches Mercury) mmHg (millimeters Mercury) InWc 4°C, 20°C, 60°F (inches of water column) cmWC 4°C, 20°C, 60°F (cm of water column) One user-defined unit NOTE: Choice of engineering unit may be restricted by available resolution.
Accuracy Pressure Range Gauge (PSIG) 0 to 0.3 (0 to 20mBar) 0 to 1 (0 to 67mBar) 0 to 5 (0 to 350 mBar) 0 to 7.2 (0 to 500 mBar) 0 to 10 (0 to 700 mBar) 0 to 30 (0 to 2 Bar) 0 to 50 (0 to 3.5 Bar) 0 to 100 (0 to 7 Bar) 0 to 150 (0 to 10 Bar) 0 to 300 (0 to 20 Bar) 0 to 500 (0 to 34 Bar) 0 to 1000 (0 to 70 Bar) 0 to 1500 (0 to 100 Bar) 0 to 3000 (0 to 200 Bar) 0 to 5000 (0 to 340 Bar) 0 to 10000 (0 to 700 Bar) Accuracy Overpressure ±0.1% ±0.05% ±0.025% ±0.035% ±0.025% ±0.025% ±0.03% ±0.025% ±0.035% ±0.
Notes: 1. Accuracy is percent of full scale range, over the 15 °C to 35 °C temperature range. Includes the pressure/temperature hysteresis in psi. The accuracy statement shown in the specification table is the base accuracy from 15 °C to 35 °C. Outside this temperature range, add an additional ±0.0015% of FS per °C. (For the 0.3 and 1psi ranges add an additional +0.
Environmental Temperature -20ºC to 50ºC (operating) 0ºC to 35ºC (storage with battery packs) -30ºC to 70ºC (storage without battery packs) Relative Humidity 0 to 95% non-condensing Effects Pressure: ± 0.0015% Range/ºC outside 15ºC to 35ºC Electrical: ±0.
Display 128 x 128 pixel LCD with LED backlight; two inputs simultaneously displayed with a maximum of 5 digits resolution Power Supply Rechargeable NiCad external battery pack: 6V, 1.
*** Specifications B-8
C- Conversion Factors Overview The following table contains the conversion factors for converting pounds per square inch (PSI) to other pressure units, i.e., multiply pressure values in PSI by the following numbers to obtain the pressure values in the corresponding units. Pressure In (Conversion Factor) x (Pressure In PSI) atmospheres (physical)* 0.068046 bar 0.068947 cm Hg (0°C) 5.17149 cm Hg (60°C) 5.1861 dynes/cm 2 68,947.57 feet H 20 (4°C) 2.306724 feet H20 (20°C) 2.
Pressure In (Conversion Factor) x (Pressure In PSI) meters sea water 0.684482 millibar 68.94757 mm Hg (0°C) 51.7149 2 Newtons/m 6,894.757 Pascals 6,894.757 Torr 51.7149 *NOTE: 1 atmosphere (physical) = 14.
D-Service Information Service Schedule Martel recommends that you return the following equipment for re-calibration and/or re-certification as follows: Equipment Service Schedule Calibrators Not required DC Modules Once yearly Pressure Modules Once yearly NOTE: Martel calibrates and certifies according to precision standards that are traceable to the National Institute of Standards and Technology.
Return Information All equipment returns should include the RMA form obtained from our web site. It has full instructions about how and where to return equipment for annual recertification or repairs either warranty or nonwarranty. Note: for warranty repairs a dated proof of purchase may be required. To obtain an RMA form, please visit our web site: www.martelcalibrators.com and click on “Customer Service.” You may also call at either 800-821-0023 or 1-603-434-1433 to have a form faxed to you.
Limited Warranty Martel Electronics Corporation warrants this product to be free from all latent defects in material and workmanship under normal use and service.
*** Service Information D-4
Index A Application Notes Overview, 6-1 Automatic Shut Off Delay Setting, 4-7 B Basic Operations Overview, 3-1 Programming Trip Detection, 3-15 Selecting a Channel for Specific Functions, 3-2 Selecting Pre-Programmed Display Units, 3-3 Setting the Damping Factor for Ch1/Ch2, 3-12 Viewing Module Information, 3-14 Zeroing DC Modules, 3-10 Zeroing Pressure Modules, 3-7 Battery Care/Maintenance Overview, A-1 Battery Packs, A-4 Charging Intrinsically Safe Battery Pack, A-10 Quick Charge Pack, A-9 Standard Pack,
C Calibration Test Temperature to Electrical Test, 6-38 Calibration Tests, 6-3 Electrical Indicator or Loop Test, 6-44 Electrical to Electrical Test, 6-22 Electrical to Pressure Test, 6-4 Pressure Indicator or Loop Test, 6-48 Pressure to Electrical Test, 6-10 Pressure to Pressure Test, 6-28 Temperature Indicator or Loop Test, 6-54 Calibrator Entering Owner Identification, 4-2 Registration, D-4 Viewing Calibrator Information, 4-3 Carrying Case, 1-2 Channel Selection, 3-2 D Damping Factor, 3-12 Reset Guideli
Programming, 5-4 User-Defined Pressure Unit Programming, 5-1 Documenting Calibrator (SYS) Mode Accessing the Execute Screen, 7-18 Tag Calibration, 7-20 Accessing the Post-Test Status Screen, 7-22 Specifying Post-Test Status Information, 7-23 Accessing the Zero Screen, 7-16 Zeroing the Display Readings, 7-17 Acquiring Instrument Maintenance Software, 7-1 Back Screen Level, 7-4 Accessing, 7-10 Before You Begin, 7-2 Connecting Necessary Equipment, 7-3 Downloading Instrument Tests, 7-2 Choosing Appropriate Modu
I Input Modules Connections, 2-3 BetaPort-DC Electrical Modules, 2-3 BetaPort-P Pressure Modules, 2-5 Installation Procedure, 2-2 Installing and Removing, 2-2 Overview, 2-1 Removal Procedure, 2-3 Viewing Module Information, 3-14 Zeroing, 3-6 DC Modules, 3-10 Pressure Modules, 3-7 K Keypad Beep Turning On/Off, 4-3 Keypad Description, 1-5 Channel Keys, 1-6 Function Keys, 1-6 Number Keys, 1-6 Special Keys, 1-7 L LCD, 1-4 Recovering the LCD Contrast, 4-6 Setting the Backlight Off Delay Time, 4-4 Setting the L
O Optional Equipment, 1-3 P Pressure Conversion Factors Overview, C-1 Pressure Module Media Compatibility, 2-6 Pressure Modules Overpressure Protection, 2-7 Safety, 2-7 R Registration Calibrator, D-4 Manual, D-4 Return Information, D-2 Packaging, D-2 Return Authorization, D-2 S Service Packaging, D-2 Return Information, D-2 Return Authorization, D-2 Service Schedule, D-1 Specifications, B-1 Accuracy, B-4 Alarm Trip Detect, B-5 Available Electrical Ranges, B-4 Available Pressure Ranges, B-1 Circuit Protec
Weight, B-6 Standard Equipment, 1-3 T Time Display Formatting, 4-9 Setting, 4-8 Trip Detection, 3-15 Dry Contacts, 3-16 Hints for Accurate Trip Measurements, 3-16 Preliminary Setup, 3-18 Procedure, 3-19 Wet Contacts, 3-15 U User Preferences Entering Owner Identification, 4-2 Formatting the Date, 4-11 Formatting the Time, 4-9 Overview, 4-1 Recovering the LCD Contrast, 4-6 Setting the Automatic Shut Off Delay Time, 4-7 Setting the Backlight Off Delay Time, 4-4 Setting the Date, 4-10 Setting the LCD Contrast
Z Zeroing Modules, 3-6 DC Modules, 3-10 Milliamp Input, 3-11 Volts Input, 3-10 Pressure Modules, 3-7 Absolute Zero Reference, 3-8 Gauge Zero Reference, 3-7 I-7 Index
Index I-8