3.6 Safety instructions
The following instructions must be observed prior to every drying process:
WARNING
Ensure that the freeze-dryer was set up and connected properly (see
chapter 5 - "Set-up and connection").
Maintain a safety distance of at least 30 cm (12 inches) around the
freeze-dryer.
Do not store any dangerous goods in the safety area of the freeze-
dryer.
Do not stay in the safety area longer than what is absolutely necessary
for the operation of the freeze-dryer.
Only use accessories that have been approved by the manufacturer
(except for commercial vessels made of glass or synthetic materials).
We explicitly warn against the use of equipment of poor quality.
Breaking glass or bursting vessels can cause dangerous situations.
Observe the instructions on the installation of accessories (see the
separate document).
Pos: 57 /010 Unive rsalm odule/ L ee rzeile @ 0\ mod _120211 624450 0_0.docx @ 11 4 @ @ 1
Pos: 58 / 200 Chris t /36 0 GT-B A La bor -Pi lot (S TA NDA RDMOD ULE) /03 0 S icher heit/ 030 -00 60- 001 0 Elekt ri sche Si cher heit @ 5\ mod _12 9913 646365 0_6 8.do cx @ 3 4716 @ 3 @ 1
3.6.1 Electrical safety
CHRIST freeze-dryers are units of safety class I. Please comply with the
following points in order to preserve the safety features:
DANGER
Ensure that the local mains voltage matches the nominal voltage that is
stated on the name plate.
Do not place any dangerous material, e.g. glass vessels containing
liquid substances, within the safety area of 30 cm around the freeze-
dryer. Spilled liquids may get into the freeze-dryer and damage the
electrical or mechanical components.
Work on the power supply system must only be performed by certified
electricians.
Inspect the electrical equipment of the unit regularly. Defects such as
loose or burnt cables must be eliminated immediately.
Pos: 59 /010 Unive rsalm odule/ Seit enwec hsel @ 0\m od_120 2116244 312_0. docx @ 105 @ @ 1