Owner manual
Gamma 1-16 LSC&
Gamma 2-16 LSC&
7 Malfunctions and error correction
74
Version 04/2013, Rev. 1.1 of 16/12/2013 • sb
Translation of the original operating manual
Pos: 158 /20 0 Chri st/3 60 GT-B A L abor-P ilot (STAN DAR DMODU LE )/0 70 Stö rungen und Fe hlers uche /07 0 Stö rungen und Fehl ersuc he== == ===== ===== ==== === ==== == @ 6\ mod _129913 751 409 1_6 8.doc x @ 3511 2 @ 1 @ 1
7 Malfunctions and error correction
Malfunctions are displayed in the dialog box "Process & equipment
messages" (see chapter 7.2 - "Process and error messages"). An
acoustic signal sounds when an error message is generated.
• Eliminate the source of the problem (see the following chapter).
• Acknowledge the error message.
Pos: 159 /01 0 Univ ersal module/ Lee rzeile @ 0\ mo d_12021 162445 00_ 0.docx @ 1 14 @ @ 1
Pos: 160 /20 0 Ch rist/3 60 G T-BA L abor-P ilot (STA NDA RDMOD U LE)/0 70 St örun gen und Fehl ers uche /07 0-0010 Al lgem eine St öru nge n Alpha - G am ma @ 6\m od_ 129 913 751 4419_68 .doc x @ 3 512 0 @ 2 @ 1
7.1 General malfunctions
Type of error Possible reason Correction
No indication on the display • No power in the mains supply
(see chapter 7.1.1 - "Power
failure").
• Power cord is not plugged in.
• Fuses have tripped.
• The mains power switch is set
to off.
• Check the mains power supply
fuse.
• Plug in the power cord correctly.
• Check the on-site fuses
• Switch mains power switch ON.
The touchpanel does not react at all
or it does not react correctly
• The sensitivity of the touchpanel
is misadjusted.
• Contact the service department
(see chapter 7.3 - "Service
contact")
The password input fails • The password is not correct. • Inform the administrator.
• If you have lost the administrator
password: contact the service
department (see chapter 7.3 -
"Service contact")
Insufficient vacuum • Incorrect connection of the
small flange connection(s).
• Loosen the connection. Place
the centring ring with the inner
sealing ring in a centred manner
between the flange connections
and connect it with the clamping
ring. Ensure that the centring
ring neither slips out of place nor
gets jammed.
• Dirty or damaged lid or door
seal.
• Clean the lid or door seal and
replace it if necessary.
• The ground-in stopper of the
attached drying chamber is not
installed correctly.
• Grease the ground-in stopper
evenly and over the entire
sealing surface with vacuum
grease.
Leakage in the media drain valve • The media drain valve is soiled
with drying residues or wool
particles from cleaning cloths.
• The O-rings are worn
• Clean the media drain valve (see
chapter 8.1.1.3 - "Aeration
valve, media drain valve")and
replace it if necessary.
• Replace the O-rings.
Leakage in a rubber valve • The valve is soiled. • Check the valves individually
(see chapter 7.1.2.4 - "Rubber
valves")