Owner manual
Gamma 1-16 LSC&
Gamma 2-16 LSC&
7 Malfunctions and error correction
Version 04/2013, Rev. 1.1 of 16/12/2013 • sb
77
Translation of the original operating manual
Pos: 169 /20 0 Ch rist/3 60 G T-BA L abo r-Pil ot (S TANDA RDMOD U LE)/0 70 Stö run gen und Fe hlers uche /07 0-0 010-00 20- 0020 Bel üf tun gsven til, Me dien abla ufven ti l @ 6\ mod_12 991 375 16122_ 68.d ocx @ 351 52 @ 4 @ 1
7.1.2.2 Aeration valve, media drain valve
A malfunction of the aeration valve or the media drain valve may have
several causes. One potential source are contaminants such as product
residues within the valve.
• Switch the freeze-dryer off and disconnect the mains plug.
• Clean the valve (see chapter 8.1.1.3 - "Aeration valve, media drain
valve").
• Put the freeze-dryer into operation again.
If there is still a leakage, the freeze-dryer must be checked by qualified
specialist personnel (see chapter 7.3 - "Service contact").
Pos: 170 /01 0 Univ ersal module/ Lee rzeile @ 0\ mo d_12021 162445 00_ 0.docx @ 1 14 @ @ 1
Pos: 171 /20 0 Chri st/3 60 GT-BA Labo r-Pi lot (STAND ARD MODU LE)/0 70 Stö run gen und Fe hlers uche /07 0-0010 -00 20-0021 D ruc kste uerv entil @ 15\ mod_1 3578 991 63437_ 68.d ocx @ 790 39 @ 4 @ 1
7.1.2.3 Pressure control valve
A malfunction of the pressure control valve may have several causes.
NOTE
The inspection of the valve must be carried out by qualified specialist
personnel (see chapter 7.3 - "Service contact").
Pos: 172 /01 0 Univ ersal module/ Lee rzeile @ 0\ mo d_12021 162445 00_ 0.docx @ 1 14 @ @ 1
Pos: 173 /01 0 Univ ersal module/ Lee rzeile @ 0\ mo d_12021 162445 00_ 0.docx @ 1 14 @ @ 1
Pos: 174 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (S TANDA RDMOD U LE)/0 70 Stö run gen und Fe hlers uche /07 0-0010 -00 20- 0030 G ummiv entile @ 6\ mo d_12991 375 164 66_ 68.doc x @ 351 60 @ 4 @ 1
7.1.2.4 Rubber valves
In order to identify a leaking rubber valve, the valves must be checked
individually:
• Remove the rubber valve and seal the connection at the drying
chamber with a rubber stopper.
• Check the tightness under vacuum until the leaking valve has been
localised.
• Clean the valve or replace it if necessary.
Pos: 175 /01 0 Univ ersal module/ Lee rzeile @ 0\ mo d_12021 162445 00_ 0.docx @ 1 14 @ @ 1
Pos: 176 /20 0 Ch rist/3 60 G T-BA L abo r-Pi lot (STAN DAR DMODU LE )/0 70 Stö run gen und Fe hlers uche /07 0-0010 -00 20- 0040 V aku ummess s ond e @ 6\mo d_1 299 1375169 66_ 68. docx @ 351 68 @ 4 @ 1
7.1.2.5 Vacuum sensor
Vacuum sensors have a limited service life and can be ordered as spare
parts (e.g. VSP62/63, order no. 125468).
Functional test:
• Connect the vacuum sensor directly to the suction side connector of the
vacuum pump.
If a final pressure of at least 0.011 mbar is reached (with a vacuum pump
that has reached its operating temperature), the sensor and the vacuum
pump are OK.
Pos: 177 /01 0 Univ ersal module/ Sei tenwe chsel @ 0\ mod_12 0211624 4312_0 .docx @ 105 @ @ 1