8.1.1.1 Ice condenser chamber
Before each start-up, ensure that the ice condenser chamber is free from
water residues.
Open the media drain valve to drain off any liquid. Then, close the
valve.
If necessary, wipe the ice condenser chamber dry with a cloth.
Pos: 192 /01 0 Univ ersal module/ Lee rzeile @ 0\ mo d_12021 162445 00_ 0.docx @ 1 14 @ @ 1
Pos: 193 /20 0 Ch ris t/3 60 G T-BA L abo r-Pi lot (STA NDA RDMOD U LE)/0 80 W a rtung u nd In st andhal t ung/ 080 -0010- 001 0-0 020 Troc k nun gska mme r Labor @ 6 \mo d_1 299 1377 272 46_68. doc x @ 352 33 @ 4 @ 1
8.1.1.2 Drying chamber
The drying chamber is equipped with the LYO CHAMBER GUARD splinter
protection film.
In order to prevent the film from being damaged, only use water and
commercially available window cleaning agents as well as synthetic
sponges, soft tissues or a rubber wiper for cleaning.
Do not dry-clean any drying chambers that are equipped with LYO
CHAMBER GUARD film!
Fig. 71: This adhesive label indicates that the drying chamber is equipped
with a splinter protection film
Pos: 194 /01 0 Univ ersal module/ Lee rzeile @ 0\ mo d_12021 162445 00_ 0.docx @ 1 14 @ @ 1
Pos: 195 /20 0 Ch rist/3 60 G T-BA L abo r-Pil ot (S TANDA RDMOD U LE)/0 80 Wa rtun g und In stan dhalt ung/ 080 -00 10- 0010-0 030 B elüf tung sventi l, Medi enabl aufv entil @ 6\ mod_12 991 377 275 27_ 68.doc x @ 352 41 @ 4 @ 1
8.1.1.3 Aeration valve, media drain valve
Contaminants such as product residues may lead to an insufficient vacuum.
In this case, the aeration valve and the media drain valve must be cleaned.
Switch the freeze-dryer off and disconnect the mains plug.
Remove the valve core.
Clean the valve core and the opening with a moist cloth.
Clean the O-rings and inspect them for any damage. Damaged O-rings
must be replaced.
1 Valve opening
2 Valve core
3 O-rings
Fig. 72: Valve opening and valve core with O-rings