8 Maintenance and service
84
Version 04/2013, Rev. 1.1 of 16/12/2013 • sb
Translation of the original operating manual
Pos: 205 /20 0 Chri st/3 60 GT-BA Labo r-Pi lot (STAND ARD MODU LE)/0 80 Wa rtun g und In stan dhalt ung/ 080 -0010- 005 0 Zub ehör @ 6\m od_1299 137 729 121_68. docx @ 35 281 @ 3 @ 1
8.1.5 Accessories
CAUTION
For the care of the accessories, special safety measures must be
considered as these are measures that will ensure operational safety at
the same time.
Chemical reactions as well as stress-corrosion (combination of oscillating
pressure and chemical reaction) can affect or destroy the metal and plastic
parts. Barely detectable cracks on the surface can expand and weaken the
material without any visible signs.
Check the material regularly (at least once a month) for
cracks
visible damage of the surface
pressure marks
signs of corrosion
other changes.
Replace any damaged components immediately for your own safety.
Immediately rinse off the accessories if any liquids that may cause
corrosion come into contact with them.
Clean the accessories outside the freeze-dryer once a week or
preferably after each use.
Pos: 206 /01 0 Univ ersal module/ Lee rzeile @ 0\ mo d_12021 162445 00_ 0.docx @ 1 14 @ @ 1
Pos: 207 /01 0 Univ ersal module/ Lee rzeile @ 0\ mo d_12021 162445 00_ 0.docx @ 1 14 @ @ 1
Pos: 208 /20 0 Ch rist/3 60 G T-BA L abor-P ilot (STA NDA RDMOD U LE)/0 80 Wa rtun g u nd In stan dhalt ung/ 080 -00 20 Des inf ektion von Tr ockn ungsk am mer und Zu beh ör @ 6\ mod _12 9913 772949 6_6 8.doc x @ 3 528 9 @ 2 @ 1
8.2 Disinfection of the drying chamber and accessories
Use commercially-available disinfectants such as, for example, Incidur
,
Meliseptol
, Sagrotan
, Buraton
, or Terralin
(available at specialised
trade).
The freeze-dryers and the accessories consist of various materials. A
possible incompatibility must be considered.
Before using cleaning or decontamination agents that were not
recommended by us, contact the manufacturer to ensure that such a
procedure will not damage the freeze-dryer.
Please contact us if you have any queries (see chapter 7.3 - "Service
contact").
DANGER
If dangerous materials (e.g. infectious and pathogenic substances) are
used, the freeze-dryer and accessories must be disinfected.