User's Manual

18 (34)
5.4 Indicators and alarms
The EMMA Capnograph is equipped with an Alarm Status Indicator and an audible alarm that may be
silenced for a period of 2 minutes.
5.4.1 Alarm signals
When an alarm is triggered, an Alarm Satus Indicator in the lower right corner of the display is lit with a
steady or blinking yellow light depending on alarm priority, together with an audible alarm beep
((()))
according to the following table:
Alarm
at t = 0
Alarm Priority: Low
at t = 20
Alarm Priority: Low
t = 40, 60, 80, ...
Alarm Priority: Medium
No Breath
((()))
((())) ((()))
((())) ((())) ((()))
Low ETCO
2
n/a
High ETCO
2
Clogged Adapter
((()))
n/a
No Adapter
((()))
Zero point
adjustment
((()))