User's Manual
Table Of Contents
- About this Manual
- Product Description, Indications for Use, Contraindications, and Features
- Safety Information, Warnings, and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology (SET)
- rainbow Pulse CO-Oximetry Technology
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring (RAM) Technology
- Chapter 2: Radical-7 Descriptions
- Chapter 3: Setup
- Chapter 4: Operation
- Using the Touchscreen and Buttons
- Using Screenlock
- Using the Home Button
- Navigating the Radical-7
- About the Display View
- Accessing the Main Menu
- Navigating the Main Menu
- Parameter Settings
- Sounds
- Device Settings
- Trends
- About
- Chapter 5: Configurations
- Chapter 6: Alarms and Messages
- About Alarms
- Silencing the Alarms
- Adaptive Threshold Alarm (ATA) Feature
- 3D Alarms
- Messages
- Replace Sensor Message
- Replace Cable Message
- Replace Adhesive Sensor Message
- Incompatible Sensor Message
- Incompatible Adhesive Sensor Message
- No Adhesive Sensor Connected Message
- Interference Detected Message
- SpO2 Only Mode Message
- Low Battery Message
- Low Perfusion Index Message
- Low Signal IQ Message
- Low SpCO SIQ Message
- Low SpMet SIQ Message
- Low SpHb SIQ Message
- Speaker Failure Message
- No Cable Connected Message
- No Sensor Connected Message
- Pulse Search Message
- Sensor Initializing Message
- Sensor Off Patient Message
- Service Required Message
- Incompatible Cable Message
- Chapter 7: Troubleshooting
- Troubleshooting Measurements
- Signal Indication and Quality Indicator (SIQ)
- Dimly Lit Parameters
- Low Perfusion
- Low Signal Quality
- SpO2 Values Do Not Correlate With Clinical Assessment or Arterial Blood Gas Measurements
- Unexpected SpO2, SpCO, SpMet, or SpHb Reading
- Unexpectedly High SpCO Reading
- Difficulty Obtaining a Reading
- SpCO Reading Displays as Dashes
- Troubleshooting the Radical-7
- Troubleshooting Measurements
- Chapter 8: Specifications
- Performance
- Accuracy
- Resolution
- Electrical
- Environmental
- Physical Characteristics
- Trending
- Alarms
- Display Indicators
- Compliance
- Output Interface
- Wireless Radio (If Installed)
- Serial Interface Specifications
- Serial Interface Setup
- Analog Output and Nurse Call Specifications
- Symbols
- ISO Country Codes (FCC and EU)
- Citations
- Chapter 9: Service and Maintenance
- Appendix: Best Practices for Comparisons to Reference Measurements
- Index
Radical-7 Chapter 9: Service and Maintenance
www.masimo.com 130 Masimo
Performance Verification
Under normal operation, no internal adjustment or recalibration is required. Safety tests
and internal adjustments should be done by qualified personnel only. Safety checks should
be performed at regular intervals or in accordance with local and governmental regulations.
To test the performance of the Radical-7 following repairs or during routine maintenance,
follow the procedure outlined in this chapter. If the Radical-7 fails any of the described tests,
discontinue its use and correct the problem before returning the instrument back to the
user.
Before performing the following tests, do the following:
Place the Handheld into the Docking Station.
Connect the Docking Station to AC power and fully charge the Handheld battery.
Disconnect any patient cables or pulse oximetry probes.
Disconnect any SatShare, serial or analog output cables from the instrument.
Set the Radical-7 to Normal operating mode by going to the Main menu and the
Home Use feature to No.
Power-On Self Test
To conduct a Power-On Self Test
1. Connect the monitor to AC power and verify that the AC Power indicator is
illuminated.
2. Turn on the monitor. Within 5 seconds, all available indicators illuminate, you
hear a tone, and the Masimo logo displays.
3. The Docking Station indicator continuously illuminates and the Radical-7 begins
normal operation.
Alarm Limit Test
To conduct an Alarm Limit Test
1. Change the High SpO2 Alarm parameter to a value two points below the
currently selected value. See Alarms for SpO2 on page 61.
2. Verify that the newly set parameter is shown on the Display screen.
3. Return the parameter to its original setting.
4. Repeat steps 1 to 3 for all active parameters.
5. Reset the alarm limits again to the original settings.