User's Manual
Table Of Contents
- About this Manual
- Product Description, Indications for Use, Contraindications, and Features
- Safety Information, Warnings, and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology (SET)
- rainbow Pulse CO-Oximetry Technology
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring (RAM) Technology
- Chapter 2: Radical-7 Descriptions
- Chapter 3: Setup
- Chapter 4: Operation
- Using the Touchscreen and Buttons
- Using Screenlock
- Using the Home Button
- Navigating the Radical-7
- About the Display View
- Accessing the Main Menu
- Navigating the Main Menu
- Parameter Settings
- Sounds
- Device Settings
- Trends
- About
- Chapter 5: Configurations
- Chapter 6: Alarms and Messages
- About Alarms
- Silencing the Alarms
- Adaptive Threshold Alarm (ATA) Feature
- 3D Alarms
- Messages
- Replace Sensor Message
- Replace Cable Message
- Replace Adhesive Sensor Message
- Incompatible Sensor Message
- Incompatible Adhesive Sensor Message
- No Adhesive Sensor Connected Message
- Interference Detected Message
- SpO2 Only Mode Message
- Low Battery Message
- Low Perfusion Index Message
- Low Signal IQ Message
- Low SpCO SIQ Message
- Low SpMet SIQ Message
- Low SpHb SIQ Message
- Speaker Failure Message
- No Cable Connected Message
- No Sensor Connected Message
- Pulse Search Message
- Sensor Initializing Message
- Sensor Off Patient Message
- Service Required Message
- Incompatible Cable Message
- Chapter 7: Troubleshooting
- Troubleshooting Measurements
- Signal Indication and Quality Indicator (SIQ)
- Dimly Lit Parameters
- Low Perfusion
- Low Signal Quality
- SpO2 Values Do Not Correlate With Clinical Assessment or Arterial Blood Gas Measurements
- Unexpected SpO2, SpCO, SpMet, or SpHb Reading
- Unexpectedly High SpCO Reading
- Difficulty Obtaining a Reading
- SpCO Reading Displays as Dashes
- Troubleshooting the Radical-7
- Troubleshooting Measurements
- Chapter 8: Specifications
- Performance
- Accuracy
- Resolution
- Electrical
- Environmental
- Physical Characteristics
- Trending
- Alarms
- Display Indicators
- Compliance
- Output Interface
- Wireless Radio (If Installed)
- Serial Interface Specifications
- Serial Interface Setup
- Analog Output and Nurse Call Specifications
- Symbols
- ISO Country Codes (FCC and EU)
- Citations
- Chapter 9: Service and Maintenance
- Appendix: Best Practices for Comparisons to Reference Measurements
- Index
Radical-7 Appendix: Best Practices for Comparisons to Reference Measurements
www.masimo.com 139 Masimo
Secure sensor cable to patient’s arm with tape.
Instruct subjects to remain still without any sensor movement.
Connect sensor cable to device without pulling on sensor.
Wait until SpHb measurement is stable (one minute) before recording
SpHb values.
Note any Low SIQ message that displays during measurement.
5. Blood sampling and laboratory analysis:
Record site of blood draw.
Document exact time blood withdrawn from patient.
When taking blood sample from arterial line, remove adequate dead
space to ensure blood sample is not diluted with solution to keep arterial
line open. When IV line set-up includes stopcocks, ensure that no fluid is
infused through the line from which you are taking the blood. Draw blood
directly into the vacutainer, if possible.
Mix blood carefully after filling the vacutainer by rotating it gently at
least 10 times. (A rotation is one turn of the tube upside down and then
right side up.)
All blood samples must be analyzed on the same calibrated laboratory
hematology analyzer (Coulter, Sysmex, or equivalent) to avoid variation
induced by the use of multiple laboratory devices.