Radical-7 Safety Information, Warnings, and Cautions
www.masimo.com 19 Masimo
Caution
Electric shock hazard: Do not open the Radical-7 cover except to replace the
battery or batteries.
Caution
To protect against injury from electric shock, follow the directions below:
Avoid placing the device on surfaces with visible liquid spills.
Do not soak or immerse the device in liquids.
Use cleaning solutions sparingly.
Caution
Electrical shock and flammability hazard: Before cleaning the Radical-7, always
turn it off and disconnect the power cord from the AC power supply.
Caution
Do not under any circumstances remove the grounding conductor from the
power plug.
Caution
Do not use extension cords or adapters of any type. The power cord and plug
must be intact and undamaged.
Caution
To ensure patient electrical isolation, connect only to other equipment with
electrically isolated circuits.
Caution
Do not connect to an electrical outlet controlled by a wall switch or dimmer.
Note: All external instrument connections to the Analog Output/Nurse Call connector must
be IEC-60950 compliant.
Note: It is recommended that the Radical-7 Handheld is docked to the Docking Station that
is attached to an AC power source when it is not in use to ensure that the battery remains
fully charged.
Note: External instrument connections to the SatShare port must be IEC-60601-1
compliant.
Note: Only use a SatShare cable that has a ferrite bead installed.