Chapter 3: Setup
The following chapter contains information about setting up the Radical-7 before use.
Unpacking and Inspection
To unpack and inspect the device
1. Remove the instrument from the shipping carton and examine it for signs of
shipping damage.
2. Check all materials against the packing list. Save all packing materials, invoice
and bill of lading. These may be required to process a claim with the carrier.
3. If anything is missing or damaged, contact the Technical Service Department.
See Return Procedure on page 132.
Docking Station Power Requirements
Always use a hospital-grade, AC power cable to connect the Docking Station to an
AC power source.
Do not connect the Docking Station to an AC outlet that is controlled by a switch
because the power to the Docking Station may be inadvertently switched off.
Verify the AC power voltage and line frequency before use.
Verify that the power source can provide an adequate power rating as indicated
on the rear panel of the Docking Station.
The Radical-7 is designed to operate on 100 to 240VAC, 47-63 Hz.
The Radical-7 is rated at 55 VA max.
Connect a hospital-grade power cable (IEC-320 connector type at the
instrument) to the Power Entry module on the Docking Station.
Connect the power cable to an AC power source.
Ensure that the instrument is adequately powered by verifying that the AC power
indicator on the Docking Station is illuminated.
See Safety Information, Warnings, and Cautions on page 11.