User's Manual
Table Of Contents
- About this Manual
- Product Description, Indications for Use, Contraindications, and Features
- Safety Information, Warnings, and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology (SET)
- rainbow Pulse CO-Oximetry Technology
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring (RAM) Technology
- Chapter 2: Radical-7 Descriptions
- Chapter 3: Setup
- Chapter 4: Operation
- Using the Touchscreen and Buttons
- Using Screenlock
- Using the Home Button
- Navigating the Radical-7
- About the Display View
- Accessing the Main Menu
- Navigating the Main Menu
- Parameter Settings
- Sounds
- Device Settings
- Trends
- About
- Chapter 5: Configurations
- Chapter 6: Alarms and Messages
- About Alarms
- Silencing the Alarms
- Adaptive Threshold Alarm (ATA) Feature
- 3D Alarms
- Messages
- Replace Sensor Message
- Replace Cable Message
- Replace Adhesive Sensor Message
- Incompatible Sensor Message
- Incompatible Adhesive Sensor Message
- No Adhesive Sensor Connected Message
- Interference Detected Message
- SpO2 Only Mode Message
- Low Battery Message
- Low Perfusion Index Message
- Low Signal IQ Message
- Low SpCO SIQ Message
- Low SpMet SIQ Message
- Low SpHb SIQ Message
- Speaker Failure Message
- No Cable Connected Message
- No Sensor Connected Message
- Pulse Search Message
- Sensor Initializing Message
- Sensor Off Patient Message
- Service Required Message
- Incompatible Cable Message
- Chapter 7: Troubleshooting
- Troubleshooting Measurements
- Signal Indication and Quality Indicator (SIQ)
- Dimly Lit Parameters
- Low Perfusion
- Low Signal Quality
- SpO2 Values Do Not Correlate With Clinical Assessment or Arterial Blood Gas Measurements
- Unexpected SpO2, SpCO, SpMet, or SpHb Reading
- Unexpectedly High SpCO Reading
- Difficulty Obtaining a Reading
- SpCO Reading Displays as Dashes
- Troubleshooting the Radical-7
- Troubleshooting Measurements
- Chapter 8: Specifications
- Performance
- Accuracy
- Resolution
- Electrical
- Environmental
- Physical Characteristics
- Trending
- Alarms
- Display Indicators
- Compliance
- Output Interface
- Wireless Radio (If Installed)
- Serial Interface Specifications
- Serial Interface Setup
- Analog Output and Nurse Call Specifications
- Symbols
- ISO Country Codes (FCC and EU)
- Citations
- Chapter 9: Service and Maintenance
- Appendix: Best Practices for Comparisons to Reference Measurements
- Index
Radical-7 Chapter 3: Setup
www.masimo.com 42 Masimo
Setting Up the Docking Station
Place the Docking Station on a stable hard flat surface near the patient. Always place the
Radical-7 on a dry surface. Maintain a minimum of 3 cm (1 inch) free space around the
Radical-7. Make sure that the Radical-7 speaker is not covered to avoid a muffled alarm
sound.
The Radical-7 Handheld, Docking Station or Standalone should not be operated outside the
following environmental conditions:
Operating Environmental Conditions
Temperature
+5°C to +40°C, +41°F to +104°F
humidity
5% to 95%, non-condensing
Operating Altitude
1060 mbar to 500 mbar pressure -1000 ft to
18,000 ft (-304 m to 5,486 m)
See Device Settings on page 74.
Initial Battery Charging
Before use, the Radical-7 Handheld battery and the Docking Station battery must be charged
completely. See Electrical Safety Information, Warnings, and Cautions on page 18.
To charge the Handheld and Docking Station for the first time
1. Attach the Handheld to the Docking Station.
2. Plug in the AC power cord to power entry module. Make sure it is securely plugged
in.
3. Plug the AC power cord into an AC power source.
4. Verify that the batteries are charging.
The Battery Charging indicators on the Docking Station flash prior to
charging and remain illuminated while the batteries are charging.
See Standalone Front Panel on page 38 and Battery Operation and Maintenance on page
127.
Setting Up for Philips, Agilent, or HP VueLink
To set up for use with VueLink compatible monitors (Philips, Agilent, or HP)
1. On the Radical-7, on the device output screen, for the serial option, select Hp
VueLink.
2. Connect one end of the VueLink cable to the Serial Output connector on the
Docking Station.
3. Connect the other end of the VueLink cable to the VueLink module and insert the
module into the VueLink compatible monitor rack.