User's Manual
Table Of Contents
- About this Manual
- Product Description, Indications for Use, Contraindications, and Features
- Safety Information, Warnings, and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology (SET)
- rainbow Pulse CO-Oximetry Technology
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring (RAM) Technology
- Chapter 2: Radical-7 Descriptions
- Chapter 3: Setup
- Chapter 4: Operation
- Using the Touchscreen and Buttons
- Using Screenlock
- Using the Home Button
- Navigating the Radical-7
- About the Display View
- Accessing the Main Menu
- Navigating the Main Menu
- Parameter Settings
- Sounds
- Device Settings
- Trends
- About
- Chapter 5: Configurations
- Chapter 6: Alarms and Messages
- About Alarms
- Silencing the Alarms
- Adaptive Threshold Alarm (ATA) Feature
- 3D Alarms
- Messages
- Replace Sensor Message
- Replace Cable Message
- Replace Adhesive Sensor Message
- Incompatible Sensor Message
- Incompatible Adhesive Sensor Message
- No Adhesive Sensor Connected Message
- Interference Detected Message
- SpO2 Only Mode Message
- Low Battery Message
- Low Perfusion Index Message
- Low Signal IQ Message
- Low SpCO SIQ Message
- Low SpMet SIQ Message
- Low SpHb SIQ Message
- Speaker Failure Message
- No Cable Connected Message
- No Sensor Connected Message
- Pulse Search Message
- Sensor Initializing Message
- Sensor Off Patient Message
- Service Required Message
- Incompatible Cable Message
- Chapter 7: Troubleshooting
- Troubleshooting Measurements
- Signal Indication and Quality Indicator (SIQ)
- Dimly Lit Parameters
- Low Perfusion
- Low Signal Quality
- SpO2 Values Do Not Correlate With Clinical Assessment or Arterial Blood Gas Measurements
- Unexpected SpO2, SpCO, SpMet, or SpHb Reading
- Unexpectedly High SpCO Reading
- Difficulty Obtaining a Reading
- SpCO Reading Displays as Dashes
- Troubleshooting the Radical-7
- Troubleshooting Measurements
- Chapter 8: Specifications
- Performance
- Accuracy
- Resolution
- Electrical
- Environmental
- Physical Characteristics
- Trending
- Alarms
- Display Indicators
- Compliance
- Output Interface
- Wireless Radio (If Installed)
- Serial Interface Specifications
- Serial Interface Setup
- Analog Output and Nurse Call Specifications
- Symbols
- ISO Country Codes (FCC and EU)
- Citations
- Chapter 9: Service and Maintenance
- Appendix: Best Practices for Comparisons to Reference Measurements
- Index
Radical-7 Chapter 4: Operation
www.masimo.com 60 Masimo
Overview of In Vivo Adjustment
The In Vivo Adjustment feature lets clinicians manually adjust one or more clinical
parameters to match that of a corresponding laboratory reference for continuous trending.
To remind clinicians that the feature is active, an offset value displays alongside the
adjusted parameter value.
When the In Vivo Adjustment is set to Yes, the feature is active (turned on) and a positive or
a negative offset value appears, as shown in the following illustration.
In Vivo offset is set to zero for any of the following:
Cable or sensor disconnect from instrument.
Sensor goes off patient causing a sensor initialization to occur.
8 hours has elapsed since the In Vivo value was activated.
The user turns off In Vivo.
Offset value
Offset value appears and indicates that In Vivo Adjustment is active. A positive value means
that the value is increased (according to a laboratory reference value as entered by a
clinician) and a negative value means the value is decreased (according to a laboratory
reference value as entered by a clinician).
In the example below, the SpO2 value is offset (highlighted) by -1.0 and SpHb is offset by
+0.4.
The In Vivo Adjustment feature can be set to Yes or No. the factory default setting is No. If
set to Yes, the parameter value is adjusted and an offset value appears. The offset value is
set by the user.
The feature applies to any of the following parameters:
In Vivo for SpO2 on page 62.
In Vivo for SpHb on page 64.
In Vivo for SpCO on page 71.
In Vivo for SpMet on page 72.