Radical-7 Chapter 4: Operation
www.masimo.com 64 Masimo
Precision
Allows the user to set the
decimal for SpHb.
0.1
0.1, 0.5, or 1.0 (whole
numbers)
Unit of
Measure
Displays total hemoglobin
(SpHb) as g/dL (grams per
deciliter) or mmol/L
(milimoles per liter).
g/dL
mmol/L or g/dL
In Vivo for SpHb
From the In Vivo screen, change any of the following options:
Options
Description
Factory Default
Settings
User Configurable
Settings
In Vivo
Calibration
Enable/Disable
See Overview of In Vivo
Adjustment on page 60.
No
Yes or No
In Vivo
Calibration
Offset
See Overview of In Vivo
Adjustment on page 60.
0.0
SpHb (± 3 g/dL with
steps of ± 0.1 g/dL)
PR Settings
From the Additional Settings screen, you change any of the following options:
PR Alarms on page 64.
Accessing Parameter Information on page 59
PR Alarms
From the PR Alarms screen, change any of the following options:
Options
Description
Factory Default
Settings
Options
High Limit
The High Limit is upper
threshold that triggers an
alarm.
140 BPM
35 to 235 BPM, in steps
of 5 BPM
Low Limit
The Low Limit is lower
threshold that triggers an
alarm.
50 BPM
30 to 230 BPM, in
steps of 5 BPM