Radical-7 Chapter 4: Operation
www.masimo.com 68 Masimo
Overview of RRp
Respiration rate can be determined by the plethysmographic waveform (RRp) using a SpO2
sensor. Respiration rate monitoring, derived from the plethysmographic waveform,
measures a patient's respiratory rate based on amplitude changes that correspond to the
respiratory cycle.
Inaccurate measurements may be caused by:
Low arterial perfusion
Motion artifact
Low arterial oxygen saturation
Excessive ambient or environmental noise
Improper sensor placement
The device can display respiration rate derived from the plethysmographic waveform under
the following conditions:
Respiration rate parameter installed on the instrument
Active optical channel
Dual Rainbow cable is disconnected
RRp Alarms
From the Alarms screen, change any of the following options:
Options
Description
Factory Default
Configurable Options
High Limit
The High Limit is
upper threshold that
triggers an alarm.
30
5 to 69 breaths per minute,
or Off
Low Limit
The Low Limit is lower
threshold that
triggers an alarm.
6
4 to 68 breaths per minute,
or Off
Alarm Delay
When an alarm
condition is met, this
feature delays the
audible part of an
alarm.
30
0, 10, 15, 30, 45