Radical-7 Chapter 4: Operation
www.masimo.com 72 Masimo
In Vivo for SpMet
From the In Vivo screen, access the following screens:
Options
Description
Factory
Default
Settings
User Configurable Settings
Enabled
Match the corresponding
laboratory reference for
continuous trending.
No
Yes or No
Offset Amount
Helps offset individual
patient bias that is
expected when comparing a
noninvasive measurement
to a laboratory reference.
---
±3.0% in steps of ± 0.1%
SpOC Settings
From the SpOC Settings screen, access the following screens:
SpOC Alarms on page 72.
Accessing Parameter Information on page 59.
SpOC Alarms
From the SpOC Alarms screen, access the following screens:
Options
Description
Factory
Default
Settings
User Configurable Settings
High Limit
The High Limit is the upper
threshold that triggers an
alarm.
Off
2% to 34% in steps of 1, or Off
Low Limit
The Low Limit is the lower
threshold that triggers an
alarm.
Off
Off, or 1% to 33% in steps of 1