User's Manual
Table Of Contents
- About this Manual
- Product Description, Indications for Use, Contraindications, and Features
- Safety Information, Warnings, and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology (SET)
- rainbow Pulse CO-Oximetry Technology
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring (RAM) Technology
- Chapter 2: Radical-7 Descriptions
- Chapter 3: Setup
- Chapter 4: Operation
- Using the Touchscreen and Buttons
- Using Screenlock
- Using the Home Button
- Navigating the Radical-7
- About the Display View
- Accessing the Main Menu
- Navigating the Main Menu
- Parameter Settings
- Sounds
- Device Settings
- Trends
- About
- Chapter 5: Configurations
- Chapter 6: Alarms and Messages
- About Alarms
- Silencing the Alarms
- Adaptive Threshold Alarm (ATA) Feature
- 3D Alarms
- Messages
- Replace Sensor Message
- Replace Cable Message
- Replace Adhesive Sensor Message
- Incompatible Sensor Message
- Incompatible Adhesive Sensor Message
- No Adhesive Sensor Connected Message
- Interference Detected Message
- SpO2 Only Mode Message
- Low Battery Message
- Low Perfusion Index Message
- Low Signal IQ Message
- Low SpCO SIQ Message
- Low SpMet SIQ Message
- Low SpHb SIQ Message
- Speaker Failure Message
- No Cable Connected Message
- No Sensor Connected Message
- Pulse Search Message
- Sensor Initializing Message
- Sensor Off Patient Message
- Service Required Message
- Incompatible Cable Message
- Chapter 7: Troubleshooting
- Troubleshooting Measurements
- Signal Indication and Quality Indicator (SIQ)
- Dimly Lit Parameters
- Low Perfusion
- Low Signal Quality
- SpO2 Values Do Not Correlate With Clinical Assessment or Arterial Blood Gas Measurements
- Unexpected SpO2, SpCO, SpMet, or SpHb Reading
- Unexpectedly High SpCO Reading
- Difficulty Obtaining a Reading
- SpCO Reading Displays as Dashes
- Troubleshooting the Radical-7
- Troubleshooting Measurements
- Chapter 8: Specifications
- Performance
- Accuracy
- Resolution
- Electrical
- Environmental
- Physical Characteristics
- Trending
- Alarms
- Display Indicators
- Compliance
- Output Interface
- Wireless Radio (If Installed)
- Serial Interface Specifications
- Serial Interface Setup
- Analog Output and Nurse Call Specifications
- Symbols
- ISO Country Codes (FCC and EU)
- Citations
- Chapter 9: Service and Maintenance
- Appendix: Best Practices for Comparisons to Reference Measurements
- Index
Radical-7 Chapter 4: Operation
www.masimo.com 85 Masimo
Changing Trend Settings
Change the maximum value and the minimum value of the Y axis for any of the available
parameters.
To adjust the trend settings for any of the available parameters
1. From the Main Menu screen, tap the Trends icon.
2. From the Trends screen, tap any of the available parameters.
3.
From the Parameter Trends screen, tap the slider for the Y-axis maximum or the
Y-axis minimum.The following is an example of the SpO2 Trend screen.