User's Manual

Rad-87 Pulse CO-Oximeter Operator’s Manual
3-3
3
Setup
System Interface Setup
Philips VueLink Setup
(Also refer to the Masimo SET VueLink Module and Serial Cable Directions for Use for additional details)
1. Select the Philips VueLink selection (PHL) from the Serial Output menu on the Rad-87.
After selecting, choose the preferred settings by stepping through menu options.
2. Connect one end of the VueLink cable to the Serial Output connector on the back of the Rad-87.
3. Insert the VueLink module into the monitor rack.
4. Disconnect the VueLink cable from the VueLink Module, if it is already connected, before
proceeding.
5. Press the VueLink button on the VueLink Module and verify that “Open Interface” is the
selected device on the HP/Agilent/Philips monitor screen.
6. Connect the VueLink cable to the VueLink Module and power on the Rad-87.
7. The SpO
2
and pulse rate values will automatically display on the Philips/Agilent/HP
monitor. The Philips/Agilent/HP monitor may be configured to display PI (Perfusion Index),
as well as optional Rainbow parameters if they are enabled in the Rad-87.
8. In order for the pleth waveform to be displayed on the Philips/Agilent/HP monitor, the user
may need to configure the Philips/Agilent/HP monitor. Please see the Philips/Agilent/HP
monitor's Operator’s Manual for complete instructions.
9. The Rad-87 Pulse CO-Oximeter can be set up to audibly indicate all patient alarms while
communicating with the Philips/ VueLink interface. Use the Interface Alarms setting in the
Output menu to enable and disable audible alarms on the Rad-87.
NOTE: During VueLink operation, audible alarms will NOT be active in the bedside
multiparameter monitor. If however, the Philips/Agilent/HP monitor is connected to a
central monitoring system, audible alarms will be available at the central monitor. The
Philips/Agilent/HP monitor may need to be configured to transfer alarms to the central
monitor. Refer to the Philips, Agilent or HP Operator's Manual for information.
RadNet Setup
1. Connect one end of the serial cable to the Serial Output connector on the back of the Rad-87.
2. Connect the other end of the serial cable to the RadNet Interface Module connector.
3. Turn the RadNet Interface Module on.
4. Select the ASCII 2 selection from the Serial options on the Rad-87.
5. A proper connection is shown by the RadNet Interface Module's Online LED being solid.
6. With a properly configured RadNet Interface Module, the Rad-87 will automatically display
the SpO
2
, PI and Pulse Rate parameters/measurements on the screen at the RadNet
Central Station.
7. The Rad-87 Pulse CO-Oximeter can be set up to audibly indicate all patient alarms while
communicating with the RadNet Interface module.
Patient SafetyNet Setup
1. Select the ASCII 2 selection from the Serial options on the Rad-87.
2. Contact Masimo installation personnel for proper installation guidance.