User's Manual

Rad-87 Pulse CO-Oximeter Operator’s Manual
4-1
4
Operation
Introduction
To operate the Rad-87 system effectively, the instrument must be set up correctly and the
operator must:
Know how the Rad-87 derives its readings (see Section 1).
Be familiar with its controls, components and operation.
Understand its status and alarm messages (see Section 5, Alarm and Messages and
Section 6, Troubleshooting).
Basic Operation
General Setup and Use
1. Inspect the Rad-87 case for damage. If damaged, refer to Section 9, Service and Repair.
2. Connect a patient cable, Dual Rainbow Cable or a direct connect sensor to the Rad-87
device. Make sure it is a firm connection and the cable is not twisted, sliced or frayed.
3. If utilizing a patient cable or Dual Rainbow Cable, select a sensor that is compatible with the
Rad-87 and the patient before connecting it to the cable. See section 8, Sensors and Patient
Cables. If using a single patient adhesive or disposable sensor, check that the emitter (red
light) and the detector are properly aligned. Remove any substances that may interfere with
the transmission of light between the sensor’s light source and detector. If using an Acoustic
Respiration Sensor, ensure that the patient's neck area is clean and dry.
4. Follow the Indications, Instructions and Cautions in the sensor's Direction for Use when
attaching, reattaching or disconnecting the sensor(s).
5. Press the Power button to turn the Rad-87 on.
6. Verify all front-panel indicators momentarily illuminate, a tone is heard, and the LED and
LCD Displays show Sensor Time remaining (if applicable).
7. Verify the front-panel display is free of alarm and system failure messages (see Section 5,
Alarms and Messages).
8. Verify that the displays shows the following (see Setup Menu Level 1; Parameter/
Measurement Alarm Limits - Screen 1, Parameter/Measurement Alarm Limits - Screen
2, Parameter/Measurement Alarm Limits - Screen 3, and Setup Menu Level 3; Set Mode
located in this chapter):
LED:
Mode setting: Standard (Std), Sleep (SLP), or Home (Hnn)
LCD:
Mode setting: Standard, Sleep, or Home
Pulse Rate Low Alarm Limit and Pulse Rate High Alarm Limit
RespirationRate (RRa) Low and High Alarm Limit
SpCO Low Alarm Limit and SpCO High Alarm Limit
SpMet Low Alarm Limit and SpMet High Alarm Limit
SpHb Low Alarm Limit and SpHb High Alarm Limit