User's Manual

4-4
Rad-87 Pulse CO-Oximeter Operator’s Manual
4
Basic Operation, continued
Default Settings
The Rad-87 Pulse CO-Oximeter stores two types of default values that the instrument
automatically retains after a power cycle.
1. Factory defaults set by Masimo.
2. Default settings that can be changed by the user will be remembered after a power cycle.
Factory Default and User Configurable Settings
OPTION FACTORY DEFAULTS USER CONFIGURABLE DEFAULTS
SpO
2
high alarm limit "---" Off 2 to 99%, then "--- "
SpO
2
low alarm limit 90% 1 to 98%
Pulse rate high alarm limit 140 BPM 35 to 235 BPM
Pulse rate low alarm limit 50 BPM 30 to 230 BPM
RRa high alarm limit (breaths per minute) 30 breaths per minute 5 to 69 breaths per minute, then "---"
RRa low alarm limit (breaths per minute) 6 breaths per minute 4 to 68 breaths per minute
SpCO high alarm limit 10
2 to
98, then "---"
SpCO low alarm limit "---" Off "---", then 1 to 97
SpMet high alarm limit 3 1 to 99.5, then "---"
SpMet low alarm limit "---" Off "---", then 0.1 to 99
SpHb high alarm limit 17 2 to 24.5, then "---"
SpHb low alarm limit 7 "---", then 1 to 24
PI high alarm limit "---" Off 0.
04 to 19, then "---"
PI low alarm limit "---" Off "---",
then 0.03 to 18
PVI high alarm limit "---" Off 2 to 99, then "---"
PVI low alarm limit "---" Off "---"
, then 1 to 98
Sensitivity APOD
Max/Normal/APOD
NOTE: MAX sensitivity will default to
APOD after a power cycle.
Display brightness Level
2 Levels 1 thru 4
Pulse tone volume Level 2 Off, Levels 1 thru 3
Alarm Silence Time 120 seconds
30, 60, 90, or 120 seconds
Alarm Volume Level 3, 70 db min Levels 1 thru 4, 87 db max
Monitoring Mode Standard (Normal) Standard, Sleep, Home
Audible Alarm Off Alarms active (On) "On/Off or muted with reminder"
Optical Sensor Off Audible Alarm Latch Off On/Off
SpO
2
Alarm Delay 5 sec 0, 5, 10, or 15 seconds
RRa Alarm Delay 30 0, 10, 15, 30, 60 seconds
Rapid Desat Alarm 5% 5, 10, Off
Serial out ASCII 2 Philips/ASCII 1/ASCII 2
Interface Alarm Alarm
Tones On Alarm Tones, Off/On
Nurse Call Type Alarm
Alarm and
Low Signal I.Q./
Low Signal I.Q./
Alarm
Nurse Call Polarity Normal Normal/Invert
Operation