User's Manual

Rad-87 Pulse CO-Oximeter Operator’s Manual
4-5
4
OPTION FACTORY DEFAULTS USER CONFIGURABLE DEFAULTS
Line Frequency 60 Hz
60 Hz, 50 Hz
SpO
2
Averaging time 8 sec 2, 4, 8, 10, 12, 14, 16
RRa Averaging Time 30 0, 10, 20, 30, 60
SmartTone Off On/Off
FastSat Off Setting is retained after power cycle
RRa Sensor Status Notifications Off On/Off
Device Profile settings N/A Blue, green, orange, pink or white
SpHb Precision 0.1 0.1/ 0.5/1
SpHb Averaging (Hb AV) Medium Long (Lng)/Medium (nnl)/Short (Sho)
SpHb Calculation (Hb Cal) Arterial Arterial (SpHb)/ Venous (SpHbv)
SpHb Display in Low SiQ No
Yes (Display SpHb value)
No (Display dashes)
PI Averaging Long Short (Sho), Long (Lng)
PVI Averaging Long Short (Sho), Long (Lng)
LCD Language English
English (Default) French, German,
Italian, Spanish, Swedish, Dutch,
Danish, Portuguese
Successful Monitoring
The following general points will aid in ensuring monitoring success.
NOTE: See Safety Information, Warnings, Cautions and Notes for additional information.
Place the sensor on a site that has sufficient perfusion and provides proper alignment of
the LED’s and detector.
Place the sensor on a site that has unrestricted blood flow.
Do not secure a sensor with tape.
Do not select a site near potential electrical interference (electro-surgical device, for
example).
Follow the Indications, Instructions and Cautions in the sensor's Direction for Use when
attaching, reattaching or disconnecting the sensor.
NOTE: When a Masimo Rainbow Sensor is properly connected to the patient, the instrument
normally goes through a 20-30 second sensor calibration/pulse search routine and
then displays numeric values installed on the instrument and supported by the sensor.
However, if the sensor calibration/pulse search routine is unsuccessful for Rainbow
parameters/measurements, the instrument automatically switches to a “SpO2 Only
Mode” to provide SpO
2
, PR, PI and PVI values.
Operation
Basic Operation, continued