User's Manual

Rad-87 Pulse CO-Oximeter Operator’s Manual
ii
Safety Information, Warnings, Cautions and Notes
The Rad-87™ Pulse CO-Oximeter™ is designed to minimize the possibility of hazards from errors in the
software program by following sound engineering design processes, Risk Analysis and Software Validation.
Explosion hazard. Do not use the Rad-87 in the presence of fl ammable anesthetics or other fl ammable
substance in combination with air, oxygen-enriched environments, or nitrous oxide.
High intensity extreme lights
(including pulsating strobe lights) directed on the sensor, may not allow the Pulse
CO-Oximeter to obtain readings.
Excessive ambient noise may affect the accuracy of the
respiration rate reading from the Acoustic Respiration
Sensor.
SpO
2
monitoring is required when monitoring RRa (Acoustic Respiration).
The Rad-87 is NOT intended for use as an apnea monitor.
The Pulse CO-Oximeter should be considered an early warning device. As a trend towards patient hypoxemia is
indicated, blood samples should be analyzed by laboratory instruments to completely understand the patient’s
condition.
Pulse rate measurement is based on the optical detection of a peripheral flow pulse and therefore may not
detect certain arrhythmias. The pulse
oximeter should not be used as a replacement or substitute for ECG
based arrhythmia analysis.
The Rad-87 is to be operated by qualifi ed personnel only. This manual, accessory directions for use, all
precautionary information, and specifi cations should be read before use.
Electric shock hazard. Do not open the Rad-87 instrument. Only a qualifi ed operator may perform maintenance
procedures specifi cally described in this manual. Refer servicing to Masimo for repair of this equipment.
Ensure that the HF surgical neutral electrode is properly connected to help prevent unintended
current return paths when using high frequency (HF) surgical equipment.
As with all medical equipment, carefully route patient cabling to reduce the possibility of patient entanglement
or strangulation.
Use cables only from the instrument manufacturer to provide protection against the effects of
discharge from a cardiac defibrillator and burns.
Do not place the Rad-87 or accessories in any position that might cause it to fall on the patient. Do not lift the
Rad-87 by the power cord or any other cable.
Interfering Substances: Dyes, or any substance containing dyes, that change usual blood pigmentation may
cause erroneous readings.
SpO
2
is empirically calibrated to functional arterial oxygen saturation in healthy adult volunteers with normal
levels of carboxyhemoglobin (COHb) and methemoglobin (MetHb). A pulse
oximeter can not measure
elevated levels of COHb or MetHb. Increases in either COHb or MetHb will affect the accuracy of the SpO
2
measurement.
For increased COHb: COHb levels above normal tend to increase the level of SpO
2
. The level of increase is
approximately equal to the amount of COHb that is present.
NOTE: High levels of COHb may occur with a seemingly normal SpO
2
. When elevated levels of COHb are
suspected, laboratory analysis (CO-Oximetry) of a blood sample should be performed.
For increased MetHb: the SpO
2
may be decreased by levels of MetHb of up to approximately 10% to 15%.
At higher levels of MetHb, the SpO
2
may tend to read in the low to mid 80s. When elevated levels of MetHb
are suspected, laboratory analysis (CO-Oximetry) of a blood sample should be performed.
Elevated levels of Methemoglobin (MetHb) will lead to inaccurate SpO
2
and SpCO
®
measurements.