User's Manual

4-8
Rad-87 Pulse CO-Oximeter Operator’s Manual
4
Message
Number
LCD
Message
Cause Condition
Message
Length
Audible
Tone
Recommendation
3†
Sensor
### Hrs
4 hours, 2
hours, 1 hour
remaining
Monitoring a
patient
120 seconds Yes
N/A
Press and hold
Display button
for 5 seconds
Displayed 3 times in display
of system settings (refer to
Section 2, LCD Display)
No
SpHb sensor
is connected
after Rad-87 is
powered on
Not monitoring
a patient
120 seconds
SpHb sensor is
applied to and/
or removed from
the patient
Monitoring a
patient
120 seconds
No
4*†
Sensor
### Min
< 1 hour
remaining
Monitoring a
patient
120 seconds Yes
Prepare
replacement SpHb
sensor
Press and hold
Display button
for 5 seconds
Displayed 3 times in display
of system settings (refer to
Section 2, LCD Display)
No
SpHb sensor
is connected
after Rad-87 is
powered on
Not monitoring
a patient
120 seconds
SpHb sensor is
applied to and/
or removed from
the patient
Monitoring a
patient
120 seconds
No
5
Replace
sensor
Sensor is non-
functional
Not monitoring
a patient
Until working sensor
attached
No
Attach working
sensor
Monitoring a
patient
Ye s
6
Incompatible
sensor
Incompatible
sensor
Monitoring or
not monitoring
a patient
Until compatible
sensor attached
No
Replace with a
compatible Masimo
Sensor (Refer to
Section 8)
Contact your
local Masimo
Representative to
learn more about
the optional SpHb
upgrade
SpHb sensor
attached to
Rad-87 without
SpHb installed
Until compatible SpHb
sensor attached
Use a non-SpHb
sensor
*While actively monitoring a patient, if the sensor time on the Rainbow reusable SpHb sensor reaches 0 minutes
remaining, the sensor will not stop monitoring the patient until the sensor is off the patient and no pulse is detected.
“###” represents the numerical value of either hours or minutes remaining on the attached Rainbow reusable SpHb
sensor.
Operation