Rad-87 Pulse CO-Oximeter Operator’s Manual
4
Operation
Setup Menu Level 2
Level 2 menu contains parameters and settings that are not changed as frequently as Level 1.
These include alarm volume, alarm delay, clear trend and button volume parameters.
Alarm Volume
BUTTONS SETTINGS
Use the Enter Button
to access the Alarm
Volume menu and to
move between Level
2 menus.
Level 3
(Default)
Use Up or Down Button to move between
settings and the Enter Button to accept
the setting and move to the next menu
screen.
OR
press the Display Button to exit without
saving the new setting and to return to
the home display screen.
Level 4
Level 1
Level 2
SpO
2
Alarm Delay
The SpO
2
alarm delay allows the user to adjust the time in which the audible status indicator will
occur after a SpO
2
alarm condition has been initiated.
BUTTONS SETTINGS
Press the Enter Button
again to move to the
next menu.
2x
5 seconds
(Default)
Use Up or Down Button to move between
settings and the Enter Button to accept the
setting and move to the next menu screen.
OR
press the Display Button to exit without
saving the new setting and to return to the
home display screen.
0 seconds
15 seconds
10 seconds