Rad-87 Pulse CO-Oximeter Operator’s Manual
4-33
4
Operation
Setup Menu Level 3, continued
SmartTone On/Off
The SmartTone feature uses a proprietary algorithm that will provide pulse tones during excessive
motion and low perfusion conditions. The pulse tone is based on an averaged pulse rate
measurement from the proprietary algorithm and may not identify irregular heart beat patterns
when there is excessive artifact present. The Normal Tone feature uses a proprietary algorithm
that will provide pulse tones during non motion and adequate perfusion conditions. In this mode,
the pulse tone may not sound if excessive artifact is present.
BUTTONS SETTING
Press the
Enter Button
again to move
to the next
menu.
Off
(Default)
(Normal Tone)
Use Up or Down Button to move between settings
AND
press the Enter Button to accept the setting and
move to the next menu.
OR
press the Display Button to exit without saving
the new setting and to return to the home display
screen.
On
(Smart Tone)
Year
BUTTONS SETTING
Press the Enter
Button again
to move to the
next menu.
Year
Use Up or Down Button to move between settings
AND
press the Enter Button to accept the setting and
move to the next menu.
OR
press the Display Button to exit without saving
the new setting and to return to the home display
screen.
Use Up or Down
Button to adjust
the setting.