Rad-87 Pulse CO-Oximeter Operator’s Manual
4
Operation
Setup Menu Level 3, continued
Interface Alarms
When Rad-87 is interfaced to another system and the Interface Alarms are set to "Alarm Tones
Off", all parameter audible alarms are muted at the Rad-87 and active at the interfaced system.
This prevents both systems from producing audible alarms at the same time.
NOTE: The Rad-87 reverts to Interface Alarms "Alarm Tones On" during power interruptions or
when the interface connection is lost. This ensures that the Rad-87 provides audible
alarms when connection to the interfaced system becomes compromised.
BUTTONS SETTING
Press the Enter
Button again to
move to the next
menu.
On
(Default)
NOTE: LCD shows
Alarm Tones
On.
Use Up or Down Button to move
between settings
AND
press the Enter Button to accept the
setting and
move to the next menu.
OR
press the Display Button to exit
without saving the new setting and to
return to the home display screen.
Off
NOTE: LCD shows
Alarm Tones
Off.