User's Manual

5-2
Rad-87 Pulse CO-Oximeter Operator’s Manual
5
Alarm Category Table, continued
CATEGORY CHANNEL ALARM
System Function
Alarms
Optical
Incompatible
Sensor
Replace
Sensor
Sen 000
Incompatible
Adhesive
Replace
Adhesive
Adh 000
Incompatible
Cable
Replace
Cable
Cbl 000
Interference
Detected
Acoustic
Incompatible
Sensor
Replace
Sensor
Sen 000
Incompatible
Cable
Replace
Cable
Cbl 000
Instrument
Error/Warning Alarms
N/A Err ## Low Battery
RRa Sensor Status
Notifications
Acoustic No Sensor
Interference
Detected
NOTE: There are no alarms associated with SpOC.
Alarm Limits
CAUTION:
• To ensure that alarm limits are appropriate for the patient being monitored, check
the limits each time the Rad-87 is used.
NOTE: The following alarm limits may be set to "---": High SpO
2
, High SpCO, Low SpCO, High
SpMet, Low SpMet, High SpHb, Low SpHb, High PI, Low PI High PVI and Low PVI.
When the setting of "---" is made on the instrument, the alarm limit is disabled.
NOTE: The low alarm limit must always be set below the high alarm setting. Attempting to
set the high alarm limit below the low alarm limit, the low alarm limit will automatically
adjust the low limit to the next setting below the newly entered high alarm limit setting.
NOTE: Pressing and holding down the up and down buttons allow for the rapid scrolling of
changing alarm limits.
NOTE: If there is a loss of power for any length of time, the Alarm settings will be set back
to the User set defaults. If the user has not utilized this option, then all parameters/
measurements will be set back to the factory defaults.
Alarms and Messages