Rad-87 Pulse CO-Oximeter Operator’s Manual
5
Alarms and Messages
MESSAGE/DISPLAY POSSIBLE CAUSE(S) RECOMMENDATION(S)
LCD: No Sensor
LED:
NO
SeN
No sensor connected. Connect sensor to patient.
LCD: Check
Sensor
LED:
CHc
SEN
Sensor is not connected
firmly into patient cable or the
instrument.
Reconnect sensor firmly into
patient cable, or to the instrument.
LCD: No Adhesive
Reconnect*
LED:
NO
AdH*
When using the Masimo
ReSposable Sensor System, the
adhesive part is not connected.
Connect the ReSposable adhesive
to the ReSposable reusable
sensor.
LCD: Replace
Adhesive*
LED:
AdH
000*
When using the Masimo
ReSposable Sensor System, the
adhesive part is non-functional.
Replace the ReSposable adhesive
of the sensor system.
LCD: Invalid
Adhesive*
LED:
IN C
ADH*
When using the Masimo
ReSposable Sensor System, the
adhesive part is incompatible or
unrecognized.
Replace the ReSposable adhesive
of the sensor system.
LCD: Replace
Adhesive*
LED:
Rpl
AdH*
When using the Masimo
ReSposable Sensor System,
the adhesive portion is non-
functional.
Replace the ReSposable adhesive
of the sensor system.
LCD: Replace
Reusable*
LED:
Rpl
REu*
When using the Masimo
ReSposable Sensor System,
the resuable sensor is non-
functional.
Replace the ReSposable
resusable sensor of the sensor
system.
LCD: Replace
Cable
LED:
rPL
CBL
The patient cable is non-
functional.
Replace the patient cable.
LCD: Replace
Cable
LED:
CBL
000
Cable is non-functional. Replace cable.
* Messages do not apply to Masimo Rainbow R Series Adhesive Sensors