User's Manual

7-2
Rad-87 Pulse CO-Oximeter Operator’s Manual
7
Respiration Rate (RRa) 1 breath per minute
Pulse Rate (bpm) 1 bpm
* Only Rainbow sensors provide ± 2% for Neonates
ELECTRICAL
AC Power requirements: 100 - 240 VAC, 47-63 Hz
Power consumption: 20 VA max.
Batteries
Type: Sealed lead acid
Capacity
(battery life)
6
:
Rad-87 (stand-alone) up to 3 hours
Rad-87 with Rainbow Acoustic Monitoring up to 2 hours
Rad-87 with wireless radio and Rainbow Acoustic Monitoring up to 1 hour
Charging time: minimum 8 hours
ENVIRONMENTAL
Operating Temperature: 32°F to 122°F (0°C to 50°C)
Transport/Storage Temperature: -40°F to 158°F (-40°C to +70°C
)7
Operating Humidity: 10% to 95%, non-condensing
Operating Altitude:
500 mbar to 1060 mbar pressure
-1000 ft to 18,000 ft (-304 m to 5,486 m)
PHYSICAL CHARACTERISTICS
Dimensions:
8.2” x 6.0” x 3.0” (20.8 cm x 15.2cm x 7.6 cm)
Weight:
2.1 lbs. = .908 Kg. = 32 oz
Trending
72 hours of trending at 2 second resolution
(Rainbow Acoustic Monitoring trending is not available)
Mode
SpO
2
Averaging mode: 2, 4, 8,10, 12, 14 or 16 seconds
8
SpO
2
Sensitivity: Normal, Maximum
9
, and APOD
RR
a Averaging mode: 0, 10, 20, 30, 60 seconds
Alarms
Audible and visual alarms for high low saturation and pulse rate (SpO
2
range 1% - 99%, SpCO range 1% - 98%,
SpMet range 1% - 99.5%, SpHb range1 g/dl - 24.5 g/dl, RRa range 4 - 70 breaths per minute, PI range 0.03% -
19%, PVI range 1% - 99%, pulse rate range 30 - 235 BPM)
Sensor condition, system failure and low battery alarms
High Priority Audible Alarm:
800 Hz tone, 5 pulse burst, pulse spacing: 0.250s,
0.250s, 0.500s, 0.250s, repeat time:10s
Medium Priority Audible Alarm 500 Hz tone, 3 pulse burst, repeat time: 5s
Low Priority Audible Alarm: 500 Hz tone, 3 pulse burst, repeat time: 5s
Alarm Volume
High:
85 dB (min)
Low: 45 dB (min)
High Priority Visual Alarm: Red flashing
2 sec. (0-5 Hz)
Medium Priority Visual Alarm Yellow flashing 4 sec (0.25 Hz)
Low Priority Visual Alarm: Solid yellow
Specifications