The Acoustic Respiration Sensor and Rainbow acoustic cables are indicated for continuous,
noninvasive monitoring of respiration rate (RRa) for adult patients, in hospitals, hospital-type
facilities, mobile and home environments.
Sensor Weight Range Breaths per Minute, Accuracy Range
RAS-125 Adult > 30 kg 4 to 70 ± 1 breath per minute
Masimo Rainbow Sensors
Rainbow sensors will only measure SpO
2
and pulse rate on devices without Masimo Rainbow
SET Technology.
Rainbow Reusable Sensors
SpO
2
, SpMet, SpHb and pulse rate accuracy for the Rainbow sensors is specified in the following
table.
Rainbow reusable sensors must be used in conjunction with Rainbow RC cables.
Sensor
Weight
Range
Saturation Accuracy
Pulse Rate
Accuracy
Low Perfusion
Accuracy
SpMet
Accuracy
SpHb
Accuracy
No Motion Motion
No
Motion
Motion Saturation
Pulse
Rate
No Motion
DCI > 30 kg
60 - 80 ± 3%
70 -100 ± 2%
± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm ± 1% ± 1 g/dl
DCIP 10 - 50 kg
60 - 80 ± 3%
70 -100 ± 2%
± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm ± 1% ± 1 g/dl
Rainbow Direct Connect Sensors
SpO
2
, SpCO, SpMet and pulse rate accuracy for the Rainbow sensors is specified in the
following table.
Rainbow Direct Connect sensors connect to the instrument directly.
Sensor
Weight
Range
Saturation Accuracy
Pulse Rate
Accuracy
Low Perfusion
Accuracy
SpCO
Accuracy
SpMet
Accuracy
No Motion Motion
No
Motion
Motion Saturation
Pulse
Rate
No Motion No Motion
DCI-dc3
DCI-dc8
DCI-dc12
> 30 kg
60 - 80 ± 3%
70 -100 ± 2%
± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm ± 3% ± 1%
DCIP-dc3
DCI-dc8
DCIP-dc12
10 - 50 kg
60 - 80 ± 3%
70 -100 ± 2%
± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm ± 3% ± 1%