Rad-87 Pulse CO-Oximeter Operator’s Manual
8
Rainbow ReSposable™ Pulse CO-Oximeter Sensor System
SpO
2
, SpMet, SpHb and pulse rate accuracy for the Rainbow sensors is specified in the following
table.
The R2-25a must be used with a R2-25r and the R2-20a must be used with a R2-20r.
Sensor
Weight
Range
Saturation
Accuracy
Pulse Rate
Accuracy
Low Perfusion Accuracy
SpMet
Accuracy
SpHb
Accuracy
Saturation Pulse Rate
R2-25a
R2-25r
> 30 kg
60 - 80% ± 3%
70 - 100% ± 2%
± 3 bpm ± 2% ± 3 bpm ± 1% ± 1 g/dl
R2-20a
R2-20r
10 - 50 kg
60 - 80% ± 3%
70 - 100% ± 2%
± 3 bpm ± 2% ± 3 bpm ± 1% ± 1 g/dl
Masimo SpO
2
Sensors
Masimo SpO2 sensors measure arterial saturation and pulse rate.
ReSposable™ Pulse CO-Oximeter Sensor System
The S2-25a must be used with a S2-25r and the S2-20a must be used with a S2-20r.
Sensor
Weight Range Saturation Accuracy
Pulse Rate
Accuracy
Low Perfusion Accuracy
Saturation Pulse Rate
S2-25a
S2-25r
> 30 kg
60 - 80% ± 3%
70 - 100% ± 2%
± 3 bpm ± 2% ± 3 bpm
S2-20a
S2-20r
10 - 50 kg
60 - 80% ± 3%
70 - 100% ± 2%
± 3 bpm ± 2% ± 3 bpm
Red Direct Connect Sensors
Red sensors will only function with CO-Oximeter devices equipped with Masimo Rainbow SET
technology.
Red Direct Connect sensors connect to the instrument directly.
Sensor
Weight Range Saturation Accuracy Pulse Rate Accuracy Low Perfusion Accuracy
No Motion Motion No Motion Motion Saturation Pulse Rate
Red DCI-dc3
Red DCI-dc12
> 30 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
Red DCIP-dc3
Red DCIP-dc12
10 - 50 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm