Operator's Manual

ENVIRONMENTAL
Storage/Transport Temperature -20°C to 50°C @ ambient humidity
Operating Temperature 10°C to 40°C @ ambient humidity
Storage/Transport Humidity 10% RH to 95% RH (non-condensing) @ ambient temperature
Operating Humidity 10% RH to 95% RH (non-condensing) @ ambient temperature
WIRELESS TECHNOLOGY INFORMATION
Type Bluetooth Low Energy
Data Transmission Rate Minimum packet rate of 0.0167 Hz (1/60 Hz)
Max. Output Power (EIRP): 9.9 dBm
Modulation Type GFSK
Frequency Range 2402–2480 MHz
Antenna Peak Gain +5.67dBi
FCC ID are as follows:
FCC ID: VKF-RADIUST, IC ID: 7362A- RADIUST
CAUTION: In order to maintain Bluetooth connectivity with the host device ensure that Radius T° is within
specied distance and line of sight of the host device.
RF Radiation Exposure Statement: This equipment has been exempted from FCC RF radiation exposure
testing and IC RSS 102 RF radiation exposure limits set forth for an uncontrolled environment.
Note: This device complies with part 15 of FCC Rules and Industry Canada’s license-exempt RSSs. Operation is
subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device
must accept any interference received, including interference that may cause undesired operation.
Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant
to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful
interference in a residential installation. This equipment generates, uses, and can radiate radio frequency
energy and, if not installed and used in accordance with the instructions, may cause harmful interference
to radio communications. However, there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to radio or television reception, which can be
determined by turning the equipment o and on, the user is encouraged to try to correct the interference by
one or more of the following measures:
Reorient or relocate the receiving antenna.
Increase the separation between the equipment and receiver.
Consult the dealer or an experienced radio/TV technician for help.
Note: When using Radius T° consideration should be taken to local government frequency allocations and
technical parameters to minimize the possibility of interference to/from other wireless devices.
RECOMMENDED SEPARATION DISTANCES
RECOMMENDED SEPARATION DISTANCE BETWEEN PORTABLE AND
MOBILE RF COMMUNICATION EQUIPMENT AND THE ME EQUIPMENT
The ME Equipment is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the ME Equipment can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the ME Equipment as recommended below,
according to the maximum output power of the communication equipment.
RATED MAXIMUM OUTPUT POWER OF
TRANSMITTER W
SEPARATION DISTANCE ACCORDING TO FREQUENCY OF TRANSMITTER M
150 kHz to 80 MHz
d = 1.17*√P
80 MHz to 800 MHz
d = 1.17*√P
800 MHz a 2.5 GHz
d = 2.33*√P
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1 1.17 1.17 2.33
10 3.7 3.7 7.37
100 11.7 11.7 23.3
For transmitters rated at a maximum output power not listed above, the recommended separation distance (d) in meters (m) can
be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of
the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reection
from structures, objects and people.
GUIDANCE AND MANUFACTURER’S DECLARATION ELECTROMAGNETIC EMISSIONS
The ME Equipment is intended for use in the electromagnetic environment specied below. The customer or the user of the ME
Equipment should assure that it is used in such an environment.
EMISSION TEST COMPLIANCE ELECTROMAGNETIC ENVIRONMENT  GUIDANCE
RF Emissions
CISPR 11
Group 1
The ME Equipment must emit electromagnetic energy in order to perform its intended
function. Nearby electronic equipment may be aected.
RF Emissions
CISPR 11
Class B Suitable for use in all establishments, including domestic environments.
GUIDANCE AND MANUFACTURER’S DECLARATION  ELECTROMAGNETIC IMMUNITY
The ME Equipment is intended for use in the electromagnetic environment specied below. The customer or the user of the
ME Equipment should assure that it is used in such an environment.
IMMUNITY TEST IEC 60601 TEST LEVEL COMPLIANCE LEVEL ELECTROMAGNETIC ENVIRONMENT  GUIDANCE
Electrostatic discharge
(ESD)
IEC 61000-4-2
+8 kV contact
+15 kV air
+8 kV contact
+15 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Power frequency (50 /
60 Hz) magnetic eld.
IEC 61000-4-8
30 A/m 30 A/m
Guidance -
Power frequency magnetic elds should be
at levels characteristic of typical location in a
typical hospital environment.
Portable and mobile RF communications equipment should be used no closer to any part of the ME Equipment, including
cables, than the recommended separation distance calculated from the equation applicable to the frequency of the
transmitter.
IMMUNITY TEST IEC 60601 TEST LEVEL COMPLIANCE LEVEL RECOMMENDED SEPARATION DISTANCE
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2.5 GHz
10 V/m
80 MHz to 800 MHz
800 MHz to 2.5 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site surveya,
should be less than the compliance level in each
frequency rangeb.
Interference may occur in the vicinity of equipment
marked with the following symbol:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio,
AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the ME
Equipment is used exceeds the applicable RF compliance level above, the ME Equipment should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the ME Equipment.
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.
TEST SPECIFICATIONS FOR ENCLOSURE PORT IMMUNITY TO RF WIRELESS COMMUNICATION EQUIPMENT
TEST
FREQUENCY
BAND A
MHZ
SERVICE A MODULATION B
MAXIMUM
POWER W
DISTANCE
M
IMMUNITY TEST
LEVEL V/M
385 380-395 TETRA 400
Pulse
modulation (b)
18 Hz
1,8 0,3 27
450 430-470
GMRS 460,
FRS 460
FM (c) +/- 5 kHz
deviation 1 kHz
sine
2 0,3 28
710
704-787 LTE Band 13, 17
Pulse
modulation (b)
217 Hz
0,2 0,3 9
745
780
810
800-960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850, LTE
Band 5
Pulse
modulation (b)
18 Hz
2 0,3 28
870
930
1 720
1 700-1 990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT; LTE
Band 1, 3. 4. 35:
UMTS
Pulse
modulation (b)
217 Hz
2 0,3 28
1 845
1 970
2 450 2 400-2 570
Bluetooth,
WLAN, 802.11
b/g/n, RFID
2450, LTE
Band 7
Pulse
modulation (b)
217 Hz
2 0,3 28
5 240
5 100-5 800
WLAN 802.11
a/n
Pulse
modulation (b)
217 Hz
0,2 0,3 9
5 500
5 785
Note: If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME
EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
(a) For some services, only the uplink frequencies are included.
(b) The carrier shall be modulated use a 50% duty cycle square wave signal.
(c) As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because while it does not represent actual
modulation, it would be worst case.
COMPATIBILITY
This sensor is intended for use only with devices containing Masimo technology.
Each sensor is designed to operate correctly only on the systems from the original
device manufacturer. Use of this sensor with other devices may result in no or
improper performance.
For Compatibility Information Reference: www. Masimo.com
WARRANTY
Masimo warrants to the initial buyer only that these products, when used in accordance with
the directions provided with the Products by Masimo, will be free of defects in materials and
workmanship for a period of six (6) months. Single use products are warranted for single
patient use only.
THE FOREGOING IS THE SOLE AND EXCLUSIVE WARRANTY APPLICABLE TO THE PRODUCTS
SOLD BY MASIMO TO BUYER. MASIMO EXPRESSLY DISCLAIMS ALL OTHER ORAL, EXPRESS
OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION ANY WARRANTIES OF
MERCHANTABILITY OR FITNESS FOR PARTICULAR PURPOSE. MASIMO’S SOLE OBLIGATION
AND BUYER’S EXCLUSIVE REMEDY FOR BREACH OF ANY WARRANTY SHALL BE, AT MASIMO’S
OPTION, TO REPAIR OR REPLACE THE PRODUCT.